Comparison of Enterprise with Neuroform stent-assisted coiling of intracranial aneurysms.

Yasha Kadkhodayan1, Nicholas Rhodes, Spiros Blackburn, Colin P Derdeyn, DeWitte T Cross, Christopher J Moran.   

Abstract

OBJECTIVE: The Enterprise stent is the first closed-cell stent designed to treat wide-necked intracranial aneurysms. Advantages of the design can include improvement in keeping coils within an aneurysm and the ability of the stent to be recaptured. We compared the technical and clinical complications of the Enterprise stent with the open-cell Neuroform stent, its primary alternative. SUBJECTS AND METHODS: Patients undergoing Enterprise and Neuroform stent-assisted aneurysm coiling were enrolled in prospective registries starting in March 2007 and February 2003, respectively. All consecutive patients through December 2011 were included. Deployment success and difficulty, stent movement and misplacement, and procedural complications were compared.
RESULTS: Enterprise deployment success was high (108 of 115 attempts, 93.9%) with 102 aneurysms receiving a stent compared with Neuroform (173 of 214 attempts, 80.8%, p = 0.001) with 163 aneurysms. Enterprise was easier to deploy (1.7% vs 15.9% difficult deployment, p < 0.0001). There were no significant differences in the rates of stent movement, misplacement, or symptomatic hemorrhage. Symptomatic thromboembolic events, however, were more frequent with the Enterprise stent (8.7% vs 1.4%, p = 0.0021). The Enterprise stent enabled treatment of 10 additional aneurysms that could not be treated with Neuroform and had a higher rate of immediate aneurysm occlusion (87.3% vs 73.0%, p = 0.0058).
CONCLUSION: Enterprise was easier to deploy and enabled treatment of additional aneurysms; however, there were more thromboembolic complications. On the basis of these findings, we prefer to use the Neuroform stent first and rely on the Enterprise stent as an easy-to-deliver backup for stent-assisted coiling.

Entities:  

Mesh:

Year:  2013        PMID: 23521463     DOI: 10.2214/AJR.12.8954

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  18 in total

1.  Stent-assisted coiling in ruptured cerebral aneurysms: multi-center experience in acute phase.

Authors:  Mario Muto; Francesco Giurazza; Gennaro Ambrosanio; Pasquale Vassallo; Francesco Briganti; Mario Tecame; Emiliano Schena; Maurizio De Nicola; Alessandro Sgreccia; Mirko Giannoni; Simone Peschillo; Francesco Diana; Giulio Guidetti; Gianluigi Guarnieri
Journal:  Radiol Med       Date:  2016-09-08       Impact factor: 3.469

2.  Comparison between Solitaire™ AB and Enterprise stent-assisted coiling for intracranial aneurysms.

Authors:  Hua-Wei Ye; Ya-Qi Liu; Qiu-Jing Wang; Tao Zheng; Xu-Bo Cui; Yu-Yuan Gao; Ling-Feng Lai; Xin Zhang; Xi-Feng Li; Shi-Xing Su; Xu-Ying He; Chuan-Zhi Duan
Journal:  Exp Ther Med       Date:  2015-05-08       Impact factor: 2.447

3.  Stent-assisted coiling of intracranial aneurysms using LEO stents: long-term follow-up in 153 patients.

Authors:  Jacques Sedat; Yves Chau; Jean Gaudart; Marina Sachet; Stephanie Beuil; Michel Lonjon
Journal:  Neuroradiology       Date:  2017-12-23       Impact factor: 2.804

4.  The impact of stent design on the structural mechanics of the crossing Y-stent: an in vitro study.

Authors:  Chang-Young Lee; Seong-Ho Park; Chang-Hyun Kim; Goetz Benndorf
Journal:  Neuroradiology       Date:  2014-05-28       Impact factor: 2.804

5.  Stent-assisted coiling of cerebral aneurysms with the Neuroform Atlas stent.

Authors:  Ferdi Cay; Ahmet Peker; Anıl Arat
Journal:  Interv Neuroradiol       Date:  2018-01-19       Impact factor: 1.610

6.  Technical failure of giant supraclinoid aneurysm after internal carotid artery occlusion. A report of three cases.

Authors:  Dong Liu; Ming Lv; Youxiang Li; Xinjian Yang; Zhongxue Wu
Journal:  Interv Neuroradiol       Date:  2014-12-05       Impact factor: 1.610

7.  Incidence of delayed ipsilateral intraparenchymal hemorrhage after stent-assisted coiling of intracranial aneurysms in a high-volume single center.

Authors:  Yasha Kayan; Josser E Delgado Almandoz; Jennifer L Fease; Kira Tran; Anna M Milner; Jill M Scholz
Journal:  Neuroradiology       Date:  2015-11-28       Impact factor: 2.804

8.  First Experiences with the New Enterprise2® Stent.

Authors:  C Herweh; S Nagel; J Pfaff; C Ulfert; M Wolf; M Bendszus; M Möhlenbruch
Journal:  Clin Neuroradiol       Date:  2016-09-27       Impact factor: 3.649

9.  Coil embolization of intracranial saccular aneurysms using the Low-profile Visualized Intraluminal Support (LVIS™) device.

Authors:  Young Dae Cho; Chul-Ho Sohn; Hyun-Seung Kang; Jeong Eun Kim; Won-Sang Cho; Gyojun Hwang; O-Ki Kwon; Mi-Sun Ko; Nam-Mi Park; Moon Hee Han
Journal:  Neuroradiology       Date:  2014-04-17       Impact factor: 2.804

Review 10.  Ocular thrombosis after stent-assisted coiling of a c7 (paraclinoid) internal carotid artery aneurysm. A report of two cases and literature review.

Authors:  Katsunari Namba; Ayuho Higaki; Shigeru Nemoto
Journal:  Interv Neuroradiol       Date:  2014-08-28       Impact factor: 1.610