Literature DB >> 23519690

Reliability of a transnasal flexible fiberoptic in-office laryngeal biopsy.

Jacob T Cohen1, Ahmad Safadi, Dan M Fliss, Ziv Gil, Gilad Horowitz.   

Abstract

IMPORTANCE: Transnasal fiberoptic laryngoscopy (TFL) has been used to guide various in-office procedures for the past 3 decades. Publications on in-office laryngeal biopsy have concurred that this procedure is safe, feasible, and easy to perform. However, the accuracy of in-office biopsy via TFL has not yet been established. The aim of this study was to examine this issue.
OBJECTIVE: To compare pathologic results obtained via in-office TFL with those of subsequent direct laryngoscopy to assess the accuracy of TFL as a diagnostic tool.
DESIGN: Prospective cohort study.
SETTING: Tertiary reference medical center. PARTICIPANTS: One-hundred two patients with suspicious laryngeal lesions. INTERVENTION: All patients underwent in-office biopsies. MAIN OUTCOME MEASURES: All patients with malignant lesions were referred to appropriate services for treatment, and those with a diagnosis of a benign lesion or carcinoma in situ were referred for direct laryngoscopy for definitive diagnosis. The results of the pathologic testing on specimens from in-office and direct laryngoscopy were compared.
RESULTS: Adequate tissue for diagnostic purposes was obtained in 96 of 102 in-office TFL biopsies (94.1%). The biopsy results revealed invasive carcinoma in 34 patients (35.4%), carcinoma in situ in 17 patients (17.7%), and benign lesions in 45 patients (46.9%). All patients with benign lesions and carcinoma in situ were referred for biopsy of samples obtained using direct laryngoscopy, to which 57 patients agreed. The final pathologic results identified from the biopsies on direct laryngoscopy revealed that there was an underestimation of the TFL results in 30 of 91 patients (false-negative rate, 33.0%) and an overestimation in 1 patient (false-positive rate, 1.1%). The sensitivity of TFL biopsy compared with that of direct laryngoscopy biopsy was 69.2% and the specificity was 96.1%. CONCLUSIONS AND RELEVANCE: Transnasal fiberoptic laryngoscopy yielded low sensitivity in assessing suspicious lesions of the larynx. These results may indicate that direct laryngoscopy represents the definitive pathologic diagnostic procedure whenever the pathologic results of an in-office TFL procedure are interpreted as benign or as carcinoma in situ.

Entities:  

Mesh:

Year:  2013        PMID: 23519690     DOI: 10.1001/jamaoto.2013.38

Source DB:  PubMed          Journal:  JAMA Otolaryngol Head Neck Surg        ISSN: 2168-6181            Impact factor:   6.223


  11 in total

1.  Diagnostic accuracy of flexible fiberoptic laryngoscopy: experience from a tertiary health institution in Nigeria.

Authors:  Habeeb K Omokanye; Sulyman B Alabi; Saadudeen O Idris'; Samuel O Ayodele; Abdulrasheed A Nasir; Ganiyu A Salaudeen; Abdul-Rahman O Afolabi; David A Dunmade
Journal:  Eur Arch Otorhinolaryngol       Date:  2021-04-23       Impact factor: 2.503

2.  Videolaryngostroboscopy versus microlaryngoscopy for the diagnosis of benign vocal cord lesions: a prospective clinical study.

Authors:  Sevtap Akbulut; Hande Altintas; Haldun Oguz
Journal:  Eur Arch Otorhinolaryngol       Date:  2014-07-18       Impact factor: 2.503

3.  Anterior meatuscopy is more reliable than a sniff test for predicting nasal patency before transnasal endoscopy.

Authors:  Chi-Tan Hu
Journal:  Endosc Int Open       Date:  2015-10-06

4.  Safety of flexible endoscopic biopsy of the pharynx and larynx under topical anesthesia.

Authors:  David J Wellenstein; Joey K de Witt; Henrieke W Schutte; Jimmie Honings; Frank J A van den Hoogen; Henri A M Marres; Robert P Takes; Guido B van den Broek
Journal:  Eur Arch Otorhinolaryngol       Date:  2017-06-21       Impact factor: 2.503

5.  An evaluation of in-office flexible fiber-optic biopsies for laryngopharyngeal lesions.

Authors:  Francisco Lee; Kristine A Smith; Shamir Chandarana; T Wayne Matthews; J Douglas Bosch; Steven C Nakoneshny; Joseph C Dort
Journal:  J Otolaryngol Head Neck Surg       Date:  2018-05-09

6.  Flexible bronchoscopy-guided vocal cord biopsy under conscious sedation: An option to surgical biopsy in special situations.

Authors:  Ravindra M Mehta; Abhinav Singla; Pavankumar Biraris; Sheelu Srinivas; Lakshmipriya Srinivasan
Journal:  Lung India       Date:  2019 May-Jun

7.  Transoral flexible laryngoscope biopsy: Safety and accuracy.

Authors:  Nabeel Humayun Hassan; Rahila Usman; Muhammad Yousuf; Ahmad Nawaz Ahmad; Ismail Hirani
Journal:  World J Otorhinolaryngol Head Neck Surg       Date:  2018-11-17

8.  Percutaneous Ultrasound-Guided Fine-Needle Aspiration Cytology and Core-Needle Biopsy for Laryngeal and Hypopharyngeal Masses.

Authors:  Dongbin Ahn; Gil Joon Lee; Jin Ho Sohn; Jeong Eun Lee
Journal:  Korean J Radiol       Date:  2020-11-30       Impact factor: 3.500

9.  Office-Based Laryngeal Biopsy in Patients Ineligible for General Anesthesia.

Authors:  Francesco Mozzanica; Francesco Ottaviani; Daniela Ginocchio; Antonio Schindler
Journal:  Iran J Otorhinolaryngol       Date:  2020-11

10.  Transnasal Flexible Fiberoptic in-office Laryngeal Biopsies-Our Experience with 117 Patients with Suspicious Lesions.

Authors:  Jacob T Cohen; Limor Benyamini
Journal:  Rambam Maimonides Med J       Date:  2014-04-28
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.