In situ osmometry: validation and effect of sample collection technique.

Moneisha Gokhale1, Ulrike Stahl, Isabelle Jalbert.   

Abstract

PURPOSE: To compare tear film osmolarity measurements between in situ and vapor pressure osmometers. Repeatability of in situ measurements and the effect of sample collection techniques on tear film osmolarity were also evaluated.
METHODS: Osmolarity was measured in one randomly determined eye of 52 healthy participants using the in situ (TearLab Corporation, San Diego, CA) and the vapor pressure (Vapro 5520; Wescor, Inc., Logan, UT) osmometers. In a subset of 20 participants, tear osmolarity was measured twice on-eye with the in situ osmometer and was additionally determined on a sample of nonstimulated collected tears (3 μL) with both instruments.
RESULTS: Mean (SD) tear film osmolarity with the in situ osmometer was 299.2 (10.3) mOsmol/L compared with 298.4 (10) mmol/kg with the vapor pressure osmometer, which correlated moderately (r = 0.5, P < 0.05). Limits of agreement between the two instruments were -19.7 to +20.5 mOsmol/L. Using collected tears, measurements with the vapor pressure osmometer were marginally higher (mean [SD], 303.0 [11.0] vs 299.3 [8.0] mOsmol/L; P > 0.05) but correlated well with those using the in situ osmometer (r = 0.9, P < 0.05). The mean (SD) osmolarity of on-eye tears was 5.0 (6.6) mOsmol/L higher than that of collected tears, when both measurements were conducted with the in situ osmometer. This was a consistent effect because the measurements correlated well (r = 0.65, P < 0.05).The in situ osmometer showed good repeatability with a coefficient of repeatability of 9.4 mOsmol/L (r = 0.8, P < 0.05).
CONCLUSIONS: Correlation between the two instruments was better when compared on collected tear samples. Tear film osmolarity measurement is influenced by the sample collection technique with the osmolarity of on-eye tears being higher than that of collected tears. This highlights the importance of measuring tear film osmolarity directly on-eye. The in situ osmometer has good repeatability for conducting this measurement.

Entities:  

Mesh:

Year:  2013        PMID: 23518677     DOI: 10.1097/OPX.0b013e31828aaf10

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  7 in total

1.  New testing options for diagnosing and grading dry eye disease.

Authors:  Gary N Foulks; Stephen C Pflugfelder
Journal:  Am J Ophthalmol       Date:  2014-03-12       Impact factor: 5.258

2.  Structural and functional changes in corneal innervation after laser in situ keratomileusis and their relationship with dry eye.

Authors:  Cecilia Chao; Fiona Stapleton; Xiangtian Zhou; Shihao Chen; Shi Zhou; Blanka Golebiowski
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-08-11       Impact factor: 3.117

Review 3.  The Growing Need for Validated Biomarkers and Endpoints for Dry Eye Clinical Research.

Authors:  Neeta S Roy; Yi Wei; Eric Kuklinski; Penny A Asbell
Journal:  Invest Ophthalmol Vis Sci       Date:  2017-05-01       Impact factor: 4.799

4.  Randomized, masked, in vitro comparison of three commercially available tear film osmometers.

Authors:  Guillermo Rocha; Eric Gulliver; Armand Borovik; Clara C Chan
Journal:  Clin Ophthalmol       Date:  2017-01-27

5.  Validity and Reliability of a Novel Handheld Osmolarity System for Measurement of a National Institute of Standards Traceable Solution.

Authors:  Clara C Chan; Armand Borovik; Ilan Hofmann; Eric Gulliver; Guillermo Rocha
Journal:  Cornea       Date:  2018-09       Impact factor: 2.651

Review 6.  Tear film osmolarity and dry eye disease: a review of the literature.

Authors:  Richard Potvin; Sarah Makari; Christopher J Rapuano
Journal:  Clin Ophthalmol       Date:  2015-11-02

7.  Utility of Tear Osmolarity Measurement in Diagnosis of Dry Eye Disease.

Authors:  Bezhod Tashbayev; Tor Paaske Utheim; Øygunn Aass Utheim; Sten Ræder; Janicke Liaaen Jensen; Mazyar Yazdani; Neil Lagali; Valeria Vitelli; Darlene A Dartt; Xiangjun Chen
Journal:  Sci Rep       Date:  2020-03-26       Impact factor: 4.379

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.