| Literature DB >> 2351721 |
S A Fuller1, D E Low, A E Simor.
Abstract
The MicroScan system was compared with agar dilution screen plates for the detection of high-level aminoglycoside resistance in 182 enterococcal isolates. Both the frozen Gram-Positive Combo Type 2 and the freeze-dried Type 5 panels were evaluated. The specificity of both panels for the detection of streptomycin and gentamicin resistance was 100%. However, the sensitivities for the detection of gentamicin and streptomycin resistance were 84 and 31%, respectively, for the Type 2 panels and 90 and 41%, respectively, for the Type 5 panels. The sensitivities of these panels for the detection of enterococcal high-level aminoglycoside resistance are inadequate for routine use.Entities:
Mesh:
Substances:
Year: 1990 PMID: 2351721 PMCID: PMC267863 DOI: 10.1128/jcm.28.5.1051-1053.1990
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948