Su-Jin Sung1, Jung-Bo Huh, Mi-Jung Yun, Brian Myung W Chang, Chang-Mo Jeong, Young-Chan Jeon.
Abstract
PURPOSE: Autoclaves and UV sterilizers have been commonly used to prevent cross-infections between dental patients and dental instruments or materials contaminated by saliva and blood. To develop a dental sterilizer which can sterilize most materials, such as metals, rubbers, and plastics, the sterilization effect of an atmospheric pressure non-thermal air plasma device was evaluated.Entities:
Keywords: Bacteria; Cross Infections; Non-thermal Atmospheric Pressure Plasma; Sterilization
Year: 2013 PMID: 23508991 PMCID: PMC3597921 DOI: 10.4047/jap.2013.5.1.2
Source DB: PubMed Journal: J Adv Prosthodont ISSN: 2005-7806 Impact factor: 1.904
Fig. 1Schematic view of an atmospheric pressure non-thermal air plasma device. A: The frontal view, B: The rear view.
Fig. 2Schematic view of a UV sterilizer.
Specification of experimental atmospheric pressure non-thermal air plasma device
Specification of UV sterilizer
Fig. 3Inoculated material. A: Diamond bur, B: Polyvinyl siloxane.
Performance tests
Mean values and standard deviations of E. coli surviving (CFU/mL) after treatment with an atmospheric pressure non-thermal air plasma device
Numbers in parentheses are standard deviations.
Mean values and standard deviations of E. coli surviving on diamond bur and polyvinyl siloxane after treatment with an atmospheric pressure non-thermal air plasma device (LogN/N0)
Fig. 4Survival curve (CFU/mL) of E. coli after treatment with an atmospheric pressure non-thermal air plasma device as a function of treatment time.
Mean values and standard deviations of B. subtilis surviving (CFU/mL) after treatment with an atmospheric pressure non-thermal air plasma device
Numbers in parentheses are standard deviations.
Mean values and standard deviations of B. subtilis surviving on diamond bur and polyvinyl siloxane after treatment with an atmospheric pressure non-thermal air plasma device (LogN/N0)
Identical superscripted letters indicate that values are not significantly different (P>.05).
Mean values and standard deviations of E. coli surviving (CFU/mL) after treatment with an atmospheric pressure non-thermal air plasma device and UV sterilizer on diamond burs
Numbers in parentheses are standard deviations.
Fig. 5Survival curve (CFU/mL) of B. subtilis after treatment with an atmospheric pressure non-thermal air plasma device as a function of treatment time.
Mean values and standard deviations of E. coli surviving after treatment with an atmospheric pressure non-thermal air plasma device and UV sterilizer on diamond burs (LogN/N0)