Literature DB >> 23492965

The VeinViewer vascular imaging system worsens first-attempt cannulation rate for experienced nurses in infants and children with anticipated difficult intravenous access.

Peter Szmuk1, Jeffrey Steiner, Radu B Pop, Alan Farrow-Gillespie, Edward J Mascha, Daniel I Sessler.   

Abstract

BACKGROUND: The VeinViewer (Luminetx, Memphis, TN) helps identify veins by projecting an image of subcutaneous vasculature on the skin surface. We tested the primary hypothesis that VeinViewer use improves cannulation success by skilled nurses in pediatric patients with anticipated difficult IV access. A secondary goal was to evaluate the relationship between obesity and cannulation success.
METHODS: Patients aged 0 to 18 years were included. Anticipated cannulation difficulty was evaluated with the difficult IV access score. All cannulations were performed by members of the Intravenous Access Team. Patients were randomized to: (1) routine IV catheter insertion; or (2) insertion facilitated by the VeinViewer. The primary outcome was first-attempt insertion success. The proportion of successful insertions was evaluated using Cochran-Mantel-Haenszel χ(2) analysis to adjust for any imbalanced baseline variables. The effect of obesity on cannulation success was evaluated with multivariable logistic regression.
RESULTS: Two hundred ninety-nine patients (49%) were randomly assigned to VeinViewer and 301 (51%) to routine cannulation. First-attempt cannulation success was 47% in patients assigned to VeinViewer vs 62% in patients assigned to routine cannulation, with an adjusted relative "risk" (95% confidence interval), of 0.76 (0.63-0.91). The Z-statistic of -3.6 crossed the "harm" boundary (Z < -2.41), with corresponding P value of 0.0003. The trial was stopped on statistical grounds since the harm boundary for the primary outcome was crossed. There was no association between first-attempt success and the 4-level categorization of obesity after adjusted for baseline variables (P = 0.94).
CONCLUSIONS: The VeinViewer worsened first-attempt IV insertion success by skilled nurses. Surprisingly, first-attempt success for IV cannulation was not worsened by obesity.

Entities:  

Mesh:

Year:  2013        PMID: 23492965     DOI: 10.1213/ANE.0b013e31828a739e

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  9 in total

Review 1.  Utility of near-infrared light devices for pediatric peripheral intravenous cannulation: a systematic review and meta-analysis.

Authors:  Joon Min Park; Min Joung Kim; Hyeon Woo Yim; Won-Chul Lee; Hyunsuk Jeong; Na Jin Kim
Journal:  Eur J Pediatr       Date:  2016-10-26       Impact factor: 3.183

2.  Utilization of a biomedical device (VeinViewer® ) to assist with peripheral intravenous catheter (PIV) insertion for pediatric nurses.

Authors:  Heidi L McNeely; Theresa L Ream; Jodi M Thrasher; Oliwier Dziadkowiec; Tiffany J Callahan
Journal:  J Spec Pediatr Nurs       Date:  2018-02-10       Impact factor: 1.260

3.  Ultrasound or near-infrared vascular imaging to guide peripheral intravenous catheterization in children: a pragmatic randomized controlled trial.

Authors:  Sarah J Curtis; William R Craig; Erin Logue; Ben Vandermeer; Amanda Hanson; Terry Klassen
Journal:  CMAJ       Date:  2015-04-20       Impact factor: 8.262

4.  Adaptive Kinematic Control of a Robotic Venipuncture Device Based on Stereo Vision, Ultrasound, and Force Guidance.

Authors:  Max L Balter; Alvin I Chen; Timothy J Maguire; Martin L Yarmush
Journal:  IEEE Trans Ind Electron       Date:  2016-04-21       Impact factor: 8.236

5.  Peripheral intravenous cannulation with support of infrared laser vein viewing system in a pre-operation setting in pediatric patients.

Authors:  Andreas Rothbart; Peng Yu; Lutz Müller-Lobeck; Claudia D Spies; Klaus-Dieter Wernecke; Irit Nachtigall
Journal:  BMC Res Notes       Date:  2015-09-21

6.  Development of the A-DIVA Scale: A Clinical Predictive Scale to Identify Difficult Intravenous Access in Adult Patients Based on Clinical Observations.

Authors:  Fredericus H J van Loon; Lisette A P M Puijn; Saskia Houterman; Arthur R A Bouwman
Journal:  Medicine (Baltimore)       Date:  2016-04       Impact factor: 1.889

7.  Performance Evaluation of Dorsal Vein Network of Hand Imaging Using Relative Total Variation-Based Regularization for Smoothing Technique in a Miniaturized Vein Imaging System: A Pilot Study.

Authors:  Kyuseok Kim; Hyun-Woo Jeong; Youngjin Lee
Journal:  Int J Environ Res Public Health       Date:  2021-02-06       Impact factor: 3.390

8.  Initial experience with use of infrared assistance for intravenous injection of radiopharmaceuticals.

Authors:  Ashok Kumar; Meena Negi; Jyoti Khanka; Mohit Dhingra; Ranjeeta Kumari; Vandana Kumar Dhingra; Manoj Kumar Gupta
Journal:  World J Nucl Med       Date:  2020-10-02

9.  Low-Cost Bicycle Lights vs. Cold Lights for Visualizing Neonatal Veins.

Authors:  Neal J Russell; Paul Bassett; John Chang
Journal:  J Trop Pediatr       Date:  2018-06-01       Impact factor: 1.165