A virtual source model of a kilo-voltage radiotherapy device.

O Nwankwo1, S Clausen, F Schneider, F Wenz.   

Abstract

The purpose of this work is to develop a virtual source model (VSM) for the 50 kV INTRABEAM® device for Monte Carlo (MC) dose calculation. The geometry of the device was modelled in Geant4. A phase space file (PSF) was computed by simulating the interactions of monoenergetic primary electrons with the target. The PSF was approximated by computing the energy spectrum of the photons in the PSF. The variation of photon intensity, mean direction cosine and standard deviation along the axis of the source was thereafter computed. The isotropy of the source was used to approximate the properties of the source on the transverse plane. These functional approximations thereafter defined the VSM of the device. A sub-source was used to account for two kinds of photons, which were suppressed by the PSF approximation method. The intensity (relative to the main source) and emission directions of the sub-source required optimization. Optimization was achieved by the iterative adjustment of either or both parameters following MC simulation with the VSM and comparison of the calculated results to experimental data. The optimized source model was validated by comparing the calculated dose to water under several experimental setups, with reference data from the manufacturer, independent dosimetric check, and to literature results. The calculated photon energy spectra at other operating potentials (30 and 40 kV) of the device were also compared to literature data. The calculated energy spectra at all operating voltages are consistent with literature reports. The optimized sub-source has a relative intensity of 5% and an emission direction that is favoured along the axis of the source. The calculated depth dose curve for the bare probe agreed with the reference data, and the isodose lines are similar to published experimental results. Validation of the source model under a more complex experimental setup by film dosimetry agreed to within 2%/1 mm (98% pixel pass rate) of the values calculated with the VSM. We derived a VSM of the 50 kVp INTRABEAM source from a PSF. The dose predicted by the model agreed with reports in literature, reference data from the device manufacturer, and with an independent validation check. The algorithm could be used for treatment planning.

Entities:  

Mesh:

Year:  2013        PMID: 23492906     DOI: 10.1088/0031-9155/58/7/2363

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  5 in total

1.  A national dosimetry audit of intraoperative radiotherapy.

Authors:  D J Eaton; B Earner; P Faulkner; N Dancer
Journal:  Br J Radiol       Date:  2013-10-16       Impact factor: 3.039

2.  [Kyphoplasty combined with intraoperative radiotherapy (Kypho-IORT). Alternative therapy for patients with oligometastatic spinal metastases].

Authors:  F Bludau; T Reis; F Schneider; S Clausen; F Wenz; U Obertacke
Journal:  Radiologe       Date:  2015-10       Impact factor: 0.635

3.  A single-source photon source model of a linear accelerator for Monte Carlo dose calculation.

Authors:  Obioma Nwankwo; Gerhard Glatting; Frederik Wenz; Jens Fleckenstein
Journal:  PLoS One       Date:  2017-09-08       Impact factor: 3.240

4.  A novel approach for superficial intraoperative radiotherapy (IORT) using a 50 kV X-ray source: a technical and case report.

Authors:  Frank Schneider; Sven Clausen; Johannes Thölking; Frederik Wenz; Yasser Abo-Madyan
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

5.  Evaluation of the dosimetric impact of manufacturing variations for the INTRABEAM x-ray source.

Authors:  Mubin Y Shaikh; Michael C Joiner; Adrian Nalichowski; Lalith K Kumaraswamy; Jay Burmeister
Journal:  J Appl Clin Med Phys       Date:  2020-01-24       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.