Performance of cassette-based blood gas analyzers to monitor blood glucose and lactate levels in a surgical intensive care setting.

Koen Poesen1, Martine De Prins, Greet Van den Berghe, Johan Van Eldere, Florent Vanstapel.   

Abstract

BACKGROUND: With the use of a traditional blood gas analyzer (BGA) (ABL800 Radiometer) for blood glucose monitoring, tight glucose control (TGC) reduced in-hospital mortality and morbidity of surgical intensive care unit (ICU) adult and pediatric patients. Such BGAs are now superseded by cassette-based BGAs. We assessed the analytical reliability of cassette-based BGAs to monitor patient's metabolic status in an ICU setting.
METHODS: We evaluated recovery/linearity, imprecision/repeatability and relative accuracy vs. comparison methods for glucose [coupled hexokinase glucose-6-phosphate dehydrogenase (HK/G6PD) assay] and lactate (lactate dehydrogenase assay) in ICU patient samples with two cassette-based BGAs [RP500 (Siemens) and ABL90 (Radiometer)] and with the ABL800 BGA.
RESULTS: Recovery of spiked glucose up to 348 mg/dL (19.3 mmol/L) was complete for all BGAs. Repeatability of ABL800 and ABL90 was comparable with the comparison method (about 1%), but higher for RP500 (about 2.4%). All BGAs were in agreement with the comparison method, with all glucose measurements falling within preset 10% criteria suggested by Karon. Recovery of spiked lactate (up to 25 mmol/L) was incomplete at all levels. Repeatability of ABL800 and ABL90 was comparable with the comparison method (about 4%), and 5.5% for RP500. All BGAs were in agreement with the comparison method, with >98% of the lactate measurements falling within 30% biological-variation-based criteria.
CONCLUSIONS: The cassette-based BGAs quantified blood glucose and lactate levels in ICU patients within the acceptable error ranges. Cassette-based BGAs can be used for monitoring patient's metabolic status in an ICU setting.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23492571     DOI: 10.1515/cclm-2012-0848

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  3 in total

1.  Analytic evaluation of a new glucose meter system in 15 different critical care settings.

Authors:  John V Mitsios; Lori A Ashby; Doris M Haverstick; David E Bruns; Mitchell G Scott
Journal:  J Diabetes Sci Technol       Date:  2013-09-01

Review 2.  Point-of-care diagnostics for niche applications.

Authors:  Brian M Cummins; Frances S Ligler; Glenn M Walker
Journal:  Biotechnol Adv       Date:  2016-02-01       Impact factor: 14.227

3.  The effect of cardiopulmonary bypass on blood thiamine concentration and its association with post-operative lactate concentration.

Authors:  Andrea L Odelli; Adam Holyoak; Sumit Yadav; Sarah M Page; Daniel Lindsay
Journal:  J Cardiothorac Surg       Date:  2022-10-07       Impact factor: 1.522

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.