Ceramic debris in hip prosthesis: correlation between synovial fluid and joint capsule.

Dalila De Pasquale1, Susanna Stea, Alina Beraudi, Monica Montesi, Stefano Squarzoni, Aldo Toni.   

Abstract

Detection of ceramic particles in synovial fluids allows early diagnosis of ceramic damage, but there is no evidence of a relationship between ceramic debris in the articular space and in the joint capsule. The aim of the present study is to verify if the particles isolated in the synovial fluid are comparable with those stored in the capsular tissue. Twenty-one patients were enrolled. Both synovial fluid and capsular samples were collected during revision surgery and ceramic particles were isolated and analyzed by scanning electron microscopy and energy-dispersive X-ray microanalysis. It resulted a significant correlation between the samples couples (18 out of 21). This study confirms that the synovial fluid analysis can give a clear definition of the presence of particles in the joint capsule.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23489725     DOI: 10.1016/j.arth.2013.01.019