Pretension effects of the Dynesys cord on the tissue responses and screw-spacer behaviors of the lumbosacral construct with hybrid fixation.

Chen-Ying Chien1, Ching-Hong Tan, Tsai-Hsiung Lu, Shang-Chih Lin, Wen-Hsien Chuang, Ming-Chieh Chiang, Yuan-Ping Luh, Yeung-Jen Chen.   

Abstract

STUDY
DESIGN: The pretension of the Dynesys cord was varied to evaluate its effects on both tissue responses and screw-spacer behaviors by the finite-element method.
OBJECTIVE: This study aimed to provide detailed information about the motion-preserving and load-shielding mechanisms of the Dynesys screw-spacer joint. SUMMARY OF BACKGROUND DATA: Intuitively, higher cord pretension aims to ensure the occurrence of screw-spacer contact, thus making the spacer the transmitter of the vertebral loads. However, detailed investigations of the cord-pretension effects have not yet been carried out. METHODS.: Using a validated lumbosacral model, the moderately degenerative L4-L5 segment was instrumented by a static fixator and the Dynesys fixator was further used to bridge a mildly degenerative L3-L4 segment. The pre-tended cord was modeled as an elastic spring with 0- and 300-N pretensions. The disc range-of-motion, disc stress, facet force, bone-screw stress, and screw-spacer force were chosen as comparison indices. RESULTS.: At the transition and adjacent segments, the range-of-motion differences between the 2 pretensions were 7.7% and 2.0% on average, respectively. The mechanical differences at the transition and adjacent segments were 9.0% and 5.2% (disc stress) and 9.4% and 9.1% (facet force), respectively. The results indicated that the cord pretension has a minor effect on the adjacent segments in comparison with the transition segment. However, the stress at the screw hub and force of the screw-spacer contact of the 300-N pretension were increased by 33.7% and 316.5% on average than without pretension, respectively.
CONCLUSION: The moment arm from the screw-cord center to the fulcrum is significantly less than that of vertebral loads. This leads to the minor effect of increasing the cord pretension on the responses of the adjacent segments. However, the cord pretension can significantly affect both screw-spacer force and bone-screw stress. LEVEL OF EVIDENCE: 4.

Mesh:

Year:  2013        PMID: 23486410     DOI: 10.1097/BRS.0b013e318290fb2e

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  4 in total

1.  Biomechanical Comparison between Isobar and Dynamic-Transitional Optima (DTO) Hybrid Lumbar Fixators: A Lumbosacral Finite Element and Intersegmental Motion Analysis.

Authors:  Shih-Hao Chen; Chih-Kun Hsiao; Chih-Wei Wang; Hsiang-Ho Chen; Zheng-Cheng Zhong
Journal:  Biomed Res Int       Date:  2022-07-08       Impact factor: 3.246

2.  A parametric investigation on traditional and cortical bone trajectory screws for transpedicular fixation.

Authors:  Tzu-Tsao Chung; Chen-Lun Chu; Dueng-Yuan Hueng; Shang-Chih Lin
Journal:  BMC Musculoskelet Disord       Date:  2022-06-27       Impact factor: 2.562

3.  The Change of Sagittal Alignment of the Lumbar Spine after Dynesys Stabilization and Proposal of a Refinement.

Authors:  Won Man Park; Chi Heon Kim; Yoon Hyuk Kim; Chun Kee Chung; Tae-Ahn Jahng
Journal:  J Korean Neurosurg Soc       Date:  2015-07-31

4.  The influence of L4-S1 Dynesys® dynamic stabilization versus fusion on lumbar motion and its relationship with lumbar degeneration: a retrospective study.

Authors:  Chengmin Zhang; Liyuan Wang; Tianyong Hou; Lei Luo; Chen Zhao; Yibo Gan; Qiang Zhou; Pei Li
Journal:  J Orthop Surg Res       Date:  2017-06-26       Impact factor: 2.359