Comprehensive hemodynamic assessment of 368 normal St. Jude Medical mechanical mitral valve prostheses based on early postimplantation echocardiographic studies.

Lori A Blauwet1, Joseph F Malouf, Heidi M Connolly, David O Hodge, Regina M Herges, Rakesh M Suri, Fletcher A Miller.   

Abstract

BACKGROUND: Two-dimensional and Doppler-derived echocardiographic data on normal St. Jude Medical mechanical mitral valve prosthesis function have been reported but remain limited.
METHODS: Comprehensive retrospective two-dimensional and Doppler echocardiographic assessment of 368 normal St. Jude Medical mechanical mitral valve prostheses was performed early after implantation. The early postimplantation hemodynamic profiles of 98 patients were compared with profiles obtained by follow-up transthoracic echocardiography performed <13 months after implantation.
RESULTS: Using mean ± 2 SDs to define the normal distribution of values for Doppler-derived hemodynamic variables, the calculated normal ranges of values were as follows: mean gradient, 2 to 7 mm Hg; peak early mitral diastolic velocity (E velocity), 1.1 to 2.4 m/sec; time-velocity integral of the mitral valve prosthesis (TVIMVP) 20 to 50 cm; ratio of the TVIMVP to the time-velocity integral of the left ventricular outflow tract (TVILVOT), 0.9 to 2.5; pressure half-time, 35 to 99 msec; and effective orifice area, 1.12 to 3.24 cm(2). Patients with severe prosthesis-patient mismatch (ie, indexed effective orifice area ≤ 0.9 cm(2)/m(2)) had significantly higher mean gradients, E velocity, TVIMVP, and TVIMVP/TVILVOT. There was a trend for longer pressure half-times for patients with severe prosthesis-patient mismatch than for patients without severe prosthesis-patient mismatch, but none of these patients had pressure half-times > 130 msec. Among the 98 patients with follow-up transthoracic echocardiography <1 year after implantation, no significant differences were observed between early postimplantation findings and follow-up hemodynamic profiles.
CONCLUSIONS: This study establishes parameters (mean ± 2 SDs) defining the distribution of values for Doppler-derived hemodynamic data with normal St. Jude Medical mechanical mitral valve prostheses. Prostheses with hemodynamic values outside these parameters are likely dysfunctional; however, prosthesis dysfunction may be present even when hemodynamic values are within these ranges.
Copyright © 2013 American Society of Echocardiography. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23484437     DOI: 10.1016/j.echo.2013.01.015

Source DB:  PubMed          Journal:  J Am Soc Echocardiogr        ISSN: 0894-7317            Impact factor:   5.251


  2 in total

1.  What Can We Expect From MitraClip After Failed Surgical Mitral Repair?

Authors:  Donald Glower
Journal:  J Am Heart Assoc       Date:  2021-04-02       Impact factor: 5.501

2.  Impact of prosthesis-patient mismatch on early and late outcomes after mitral valve replacement: a meta-analysis.

Authors:  Meng-Wei Tan; Yi-Fan Bai; Xiao-Hong Liu; Zhi-Yun Xu; Zhao An; Ye Ma; Li-Bo Zhao; Bai-Ling Li
Journal:  J Geriatr Cardiol       Date:  2020-08       Impact factor: 3.327

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.