Abstract
Three automated immunoassay kits for anti-Hepatitis A Virus (HAV) IgM-Architect, (Abbott Laboratories, USA), Elecsys (Roche Diagnostics, Germany), and ADVIA Centaur (Siemens Healthcare Diagnostics Inc., USA)-were compared. We included 178 consecutive samples, for which an anti-HAV IgM test was requested at Seoul National University Hospital from September 2009 to January 2010. Reviewing of medical records, reverse transcription (RT)-PCR for HAV RNA, or total anti-HAV assay were performed on 16 (9.0%) samples with discrepant results. The percent agreements (kappas) of the Architect and ADVIA Centaur, Architect and Elecsys, and ADVIA Centaur and Elecsys kits were 96.6% (0.91), 96.6% (0.92), and 97.8% (0.94), respectively. Eight out of 16 discrepant samples showed gray-zone values in Architect but were nonreactive in the others. Slightly earlier seroconversion was suspected in Elecsys. The 3 assays showed comparable performances with excellent agreements in a tertiary care hospital setting.Entities:
Keywords: Agreement; Anti-hepatitis A virus immunoglobulin M; Immunoassay
Mesh:
Substances:
Year: 2013 PMID: 23483019 PMCID: PMC3589637 DOI: 10.3343/alm.2013.33.2.121
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
Clinical characteristics of cases with discrepant results among Architect, ADVIA Centaur, and Elecsys Anti-HAV IgM assays (N=16)
*For Architect HAVAb-IgM, specimens with signal-to-cutoff (S/CO) values 0.80-1.20 were considered gray-zone. For ADVIA Centaur HAV IgM, S/CO values ≥0.80 and <1.20 were considered equivocal.
Abbreviations: T.bil, total bilirubin; D.bil, direct bilirubin; F/U, follow up; HAV, hepatitis A virus; N, nonreactive or negative; G, gray zone; R, reactive; P, positive; NT, not tested; S/CO, signal-to-cutoff; RT-PCR, reverse transcription-PCR.
Fig. 1Correlations of signal-to-cutoff (S/CO) values among Architect HAVAb-IgM, ADVIA Centaur HAV IgM, and Elecsys Anti-HAV IgM assays. (A) Scatter plot of S/CO values of Architect HAVAb-IgM and ADVIA Centaur HAV IgM assays, (B) Architect HAVAb-IgM and Elecsys Anti-HAV IgM assays, and (C) ADVIA Centaur HAV IgM and Elecsys Anti-HAV IgM assays.
Abbreviation: HAV, hepatitis A virus.