Evaluation of corneal displacement using high-speed photography at the early and late phases of noncontact tonometry.

Shunsuke Nakakura1, Yoshiaki Kiuchi, Makoto Kaneko, Hideki Mochizuki, Joji Takenaka, Kenji Yamada, Yosuke Kimura, Hitoshi Tabuchi.   

Abstract

PURPOSE: We investigated the extent of corneal displacement and factors associated with the measurement of IOP during the early and late phases of noncontact tonometry.
METHODS: One eye in each of 61 healthy volunteers was studied. In each case, the cornea was photographed in profile using a high-speed camera during noncontact tonometry. The extent of displacement at the corneal center, and at the midperipheral corneal areas at 13.2 and 30.0 ms after the application of an air puff was recorded. For each measurement point, multiple regression analyses were performed against age, sex, axial length, corneal curvature, IOP, and central corneal thickness (CCT) to determine the independent predictors of corneal displacement. Multiple comparison tests were used to determine whether the displacement differences were associated with age or sex.
RESULTS: Our results showed that the factors most associated with corneal displacement were the IOP (P < 0.001) and CCT (P = 0.02) at the corneal center at 13.2 ms, the CCT (P = 0.041) at the corneal midperiphery at 13.2 ms, age (P < 0.01) at the corneal center at 30.0 ms, and age (P = 0.04) and IOP (P = 0.04) at the corneal midperiphery at 30.0 ms. Young females had the smallest amounts of corneal displacement at all measurement points, and elderly females had largest amounts of displacement at the late phase of tonometry.
CONCLUSIONS: The amount of corneal displacement is affected by an individual's IOP, age, and CCT. Our findings indicated that the corneas of females may be more susceptible to aging than those of males.

Entities:  

Mesh:

Year:  2013        PMID: 23482468     DOI: 10.1167/iovs.12-11424

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  4 in total

1.  Ocular surface displacement with and without contact lenses during non-contact tonometry.

Authors:  Ulfah Rimayanti; Yoshiaki Kiuchi; Shohei Uemura; Joji Takenaka; Hideki Mochizuki; Makoto Kaneko
Journal:  PLoS One       Date:  2014-04-29       Impact factor: 3.240

2.  A Comparison of the Corrected Intraocular Pressure Obtained by the Corvis ST and Reichert 7CR Tonometers in Glaucoma Patients.

Authors:  Yoshitaka Nakao; Yoshiaki Kiuchi; Satoshi Okimoto
Journal:  PLoS One       Date:  2017-01-17       Impact factor: 3.240

3.  Cataract surgery causes biomechanical alterations to the eye detectable by Corvis ST tonometry.

Authors:  Yoshitake Kato; Shunsuke Nakakura; Ryo Asaoka; Kanae Matsuya; Yuki Fujio; Yoshiaki Kiuchi
Journal:  PLoS One       Date:  2017-02-21       Impact factor: 3.240

4.  Morphological prediction of glaucoma by quantitative analyses of ocular shape and volume using 3-dimensional T2-weighted MR images.

Authors:  Yasuko Tatewaki; Tatsushi Mutoh; Kazuko Omodaka; Benjamin Thyreau; Izumi Matsudaira; Hiroaki Furukawa; Keiji Yamada; Keiko Kunitoki; Ryuta Kawashima; Toru Nakazawa; Yasuyuki Taki
Journal:  Sci Rep       Date:  2019-10-22       Impact factor: 4.379

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.