| Literature DB >> 23481406 |
C Margo1, J Katrib, M Nadi, A Rouane.
Abstract
This paper presents the design of a four electrode impedance measurement circuit dedicated to bioimpedance embedded applications. It extends the use of the AD5933 measurement chip to allow it to operate in a four electrode configuration in order to limit the effects of the parasitic impedances between the medium under test and the electrodes. The circuit has shown a good measurement accuracy on various test circuits. In association with a four microband electrode system it has been successfully used to characterize small physiological samples (50 μl) with conductivities ranging from 0.14 to 1.2 S m(-1). It can be used as an alternative bioimpedance measurement approach for embedded applications operating in the four electrode configuration.Mesh:
Substances:
Year: 2013 PMID: 23481406 DOI: 10.1088/0967-3334/34/4/391
Source DB: PubMed Journal: Physiol Meas ISSN: 0967-3334 Impact factor: 2.833