A four-electrode low frequency impedance spectroscopy measurement system using the AD5933 measurement chip.

C Margo1, J Katrib, M Nadi, A Rouane.   

Abstract

This paper presents the design of a four electrode impedance measurement circuit dedicated to bioimpedance embedded applications. It extends the use of the AD5933 measurement chip to allow it to operate in a four electrode configuration in order to limit the effects of the parasitic impedances between the medium under test and the electrodes. The circuit has shown a good measurement accuracy on various test circuits. In association with a four microband electrode system it has been successfully used to characterize small physiological samples (50 μl) with conductivities ranging from 0.14 to 1.2 S m(-1). It can be used as an alternative bioimpedance measurement approach for embedded applications operating in the four electrode configuration.

Mesh:

Substances:

Year:  2013        PMID: 23481406     DOI: 10.1088/0967-3334/34/4/391

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  3 in total

1.  Robust Longitudinal Ankle Edema Assessment Using Wearable Bioimpedance Spectroscopy.

Authors:  Samer Mabrouk; Sinan Hersek; Hyeon Ki Jeong; Daniel Whittingslow; Venu G Ganti; Paul Wolkoff; Omer T Inan
Journal:  IEEE Trans Biomed Eng       Date:  2019-07-10       Impact factor: 4.538

2.  POISED-5, a portable on-board electrochemical impedance spectroscopy biomarker analysis device.

Authors:  M Anne Sawhney; R S Conlan
Journal:  Biomed Microdevices       Date:  2019-07-04       Impact factor: 2.838

3.  Self-Abrading Servo Electrode Helmet for Electrical Impedance Tomography.

Authors:  James Avery; Brett Packham; Hwan Koo; Ben Hanson; David Holder
Journal:  Sensors (Basel)       Date:  2020-12-09       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.