[Faulty internal tube in a co-axial ventilation tube system: cause of a massive postoperative hypercapnia].

J-H Günther1, P Börning, L Bahlmann.   

Abstract

This article presents the case of a patient with massive postoperative hypercapnia during mechanical ventilation in the intensive care unit (ICU). With normal tidal volumes and clearly visible chest movements, adequate findings with regard to auscultation, oxygenation and correct respirator settings, no cause for the increasing hypercapnia was initially found; however, replacement of the respirator led to a return to normal carbon dioxide levels. When checking the replaced respirator a service technician found the cause of the respirator failure: the internal tube of the co-axial ventilation system was faulty leading to an increased dead space and rebreathing of carbon dioxide.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23475174     DOI: 10.1007/s00101-013-2147-0

Source DB:  PubMed          Journal:  Anaesthesist        ISSN: 0003-2417            Impact factor:   1.041


  6 in total

1.  Hypercapnia related to a faulty adult co-axial breathing circuit.

Authors:  W S Jellish; T Nolan; B Kleinman
Journal:  Anesth Analg       Date:  2001-10       Impact factor: 5.108

2.  The use of capnography and the availability of airway equipment on Intensive Care Units in the UK and the Republic of Ireland.

Authors:  Andrew P Georgiou; S Gouldson; A M Amphlett
Journal:  Anaesthesia       Date:  2010-03-19       Impact factor: 6.955

3.  Hypercarbia and arrhythmias resulting from faulty Bain circuit: a report of two cases.

Authors:  Babita Ghai; Jeetinder Kaur Makkar; Anshul Bhatia
Journal:  Anesth Analg       Date:  2006-06       Impact factor: 5.108

4.  The Bain anesthesia circuit.

Authors:  J A Bain
Journal:  Int Anesthesiol Clin       Date:  1982

5.  Hypercarbia induced by detachment of inner fresh gas delivery tube of the bain circuit--a case report.

Authors:  Z T Jouan; B Jawan; J H Lee
Journal:  Acta Anaesthesiol Sin       Date:  1995-06

6.  Hypercapnia resulting from a faulty co-axial (Bain) circuit.

Authors:  Ishwar Singh; Monika Gupta; Tarun K Singh
Journal:  Indian J Anaesth       Date:  2011-07
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.