Droplets produced by medical nebulizers. Some factors affecting their size and solute concentration.

P R Phipps1, I Gonda.   

Abstract

The effect of nebulizer solution temperature and dilution air humidity on the size and solute concentration of aqueous aerosol droplets were studied. Four combinations of jet-nebulizers with air compressors or oxygen sources and one ultrasonic nebulizer were tested. The temperature to which the nebulizer solution of each system fell during generation was measured. The nebulizers were then kept at set temperatures, generated aerosols collected and either droplet size or solute concentration measured. The droplet solute concentration was found to increase. The droplet size decreased along with the droplet solute concentration increase. The ultrasonic nebulizer also was tested: its high output made the concentration of the solution in the droplets much more stable. However, the proportion of droplets depositing in the tubing and valves changed markedly with aerosol flow rate. The potential for large changes in droplet solute concentration, droplet size and output during nebulization should be considered in therapeutic and diagnostic applications of nebulized aerosols.

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2347217     DOI: 10.1378/chest.97.6.1327

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  14 in total

Review 1.  The science of nebulised drug delivery.

Authors:  C O'Callaghan; P W Barry
Journal:  Thorax       Date:  1997-04       Impact factor: 9.139

Review 2.  Canadian Asthma Consensus Report, 1999. Canadian Asthma Consensus Group.

Authors:  L P Boulet; A Becker; D Bérubé; R Beveridge; P Ernst
Journal:  CMAJ       Date:  1999-11-30       Impact factor: 8.262

Review 3.  Aerosols for therapy and diagnosis.

Authors:  M J O'Doherty; R F Miller
Journal:  Eur J Nucl Med       Date:  1993-12

4.  Characterization of aerosols of human recombinant deoxyribonuclease I (rhDNase) generated by jet nebulizers.

Authors:  D Cipolla; I Gonda; S J Shire
Journal:  Pharm Res       Date:  1994-04       Impact factor: 4.200

5.  Nebulization of Cyclic Arginine-Glycine-(D)-Aspartic Acid-Peptide Grafted and Drug Encapsulated Liposomes for Inhibition of Acute Lung Injury.

Authors:  Hari R Desu; Laura A Thoma; George C Wood
Journal:  Pharm Res       Date:  2018-03-13       Impact factor: 4.200

6.  Aerosol kinetics and bronchodilator efficacy during continuous positive airway pressure delivered by face mask.

Authors:  S N Parkes; A D Bersten
Journal:  Thorax       Date:  1997-02       Impact factor: 9.139

7.  Bronchodilator aerosol administered by metered dose inhaler and spacer in subacute neonatal respiratory distress syndrome.

Authors:  H Lee; S Arnon; M Silverman
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1994-05       Impact factor: 5.747

8.  Inhalable curcumin: offering the potential for translation to imaging and treatment of Alzheimer's disease.

Authors:  Richard McClure; Daijiro Yanagisawa; Donald Stec; Dave Abdollahian; Dmitry Koktysh; Dritan Xhillari; Rudolph Jaeger; Gregg Stanwood; Eduard Chekmenev; Ikuo Tooyama; John C Gore; Wellington Pham
Journal:  J Alzheimers Dis       Date:  2015       Impact factor: 4.472

9.  Short-term exposure to urban PM2.5 particles induces histopathological and inflammatory changes in the rat small intestine.

Authors:  Lena Ohlsson; Christina Isaxon; Sebastian Wrighton; Wissal El Ouahidi; Lisa Fornell; Lena Uller; Saema Ansar; Ulrikke Voss
Journal:  Physiol Rep       Date:  2022-04

10.  Collecting Protein Biomarkers in Breath Using Electret Filters: A Preliminary Method on New Technical Model and Human Study.

Authors:  Wang Li; Xitian Pi; Panpan Qiao; Hongying Liu
Journal:  PLoS One       Date:  2016-03-02       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.