Comparison of the x-ray attenuation properties of breast calcifications, aluminium, hydroxyapatite and calcium oxalate.

L M Warren1, A Mackenzie, D R Dance, K C Young.   

Abstract

Aluminium is often used as a substitute material for calcifications in phantom measurements in mammography. Additionally, calcium oxalate, hydroxyapatite and aluminium are used in simulation studies. This assumes that these materials have similar attenuation properties to calcification, and this assumption is examined in this work. Sliced mastectomy samples containing calcification were imaged at ×5 magnification using a digital specimen cabinet. Images of the individual calcifications were extracted, and the diameter and contrast of each calculated. The thicknesses of aluminium required to achieve the same contrast as each calcification when imaged under the same conditions were calculated using measurements of the contrast of aluminium foils. As hydroxyapatite and calcium oxalate are also used to simulate calcifications, the equivalent aluminium thicknesses of these materials were also calculated using tabulated attenuation coefficients. On average the equivalent aluminium thickness was 0.85 times the calcification diameter. For calcium oxalate and hydroxyapatite, the equivalent aluminium thicknesses were 1.01 and 2.19 times the thickness of these materials respectively. Aluminium and calcium oxalate are suitable substitute materials for calcifications. Hydroxyapatite is much more attenuating than the calcifications and aluminium. Using solid hydroxyapatite as a substitute for calcification of the same size would lead to excessive contrast in the mammographic image.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23470559     DOI: 10.1088/0031-9155/58/7/N103

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Classification of breast microcalcifications using dual-energy mammography.

Authors:  Bahaa Ghammraoui; Andrey Makeev; Ahmed Zidan; Alaadin Alayoubi; Stephen J Glick
Journal:  J Med Imaging (Bellingham)       Date:  2019-03-12

2.  Exploring CNN potential in discriminating benign and malignant calcifications in conventional and dual-energy FFDM: simulations and experimental observations.

Authors:  Andrey Makeev; Gabriela Rodal; Bahaa Ghammraoui; Andreu Badal; Stephen J Glick
Journal:  J Med Imaging (Bellingham)       Date:  2021-05-13

3.  Phantom-Based Feasibility Studies on Phase-Contrast Mammography at Indian Synchrotron Facility Indus-2.

Authors:  Reena Sharma; S D Sharma; P S Sarkar; B Singh; A K Agrawal; D Datta
Journal:  J Med Phys       Date:  2019 Jan-Mar
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.