Study of light transport inside scintillation crystals for PET detectors.

Xin Yang1, Evan Downie, Thomas Farrell, Hao Peng.   

Abstract

Scintillation crystal design is a critical component in positron emission tomography system development, which impacts a number of performance parameters including energy resolution, time resolution and spatial resolution. Our work aims to develop a generalized simulation tool to model the light transport inside scintillation crystals with good accuracy, taking into account surface treatments, reflectors, temporal dependence of scintillation decay, and comprehensive experimental validations. The simulation has been validated against both direct analytical calculation and experimental measurements. In this work, the studies were performed for a lutetium-yttrium oxyorthosilicate crystal of 3×3×20 mm(3) dimension coupled to a Hamamatsu silicon photomultiplier, with respect to light output, rise-time slope, energy resolution and time resolution. Four crystal surface treatment and reflector configurations were investigated: GroundMetal, GroundPaint, PolishMetal and PolishPaint. The experiments were performed to validate the Monte Carlo simulation results. The results indicate that the best time resolution (0.96±0.05 ns) and good energy resolution (10.6±0.4%) could be produced by using a polished surface with specular reflector, while the configuration of a polished surface with diffusive reflector produces the best energy resolution (10.2±0.9%). The results indicate that a polished surface with diffusive reflector achieves the best energy resolution (10.2±0.9%) for 511 keV high energy photons, and a polished surface with specular reflector achieves the best time resolution (0.96±0.05 ns) measured against a Hamamatsu fast photomultiplier tube. The ground surface treatment is not recommended for its inferior performance in terms of energy and time resolution. Possible explanations and future improvements to be made to the developed simulation tool are discussed.

Entities:  

Mesh:

Year:  2013        PMID: 23470488     DOI: 10.1088/0031-9155/58/7/2143

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  Analytical calculation of the lower bound on timing resolution for PET scintillation detectors comprising high-aspect-ratio crystal elements.

Authors:  Joshua W Cates; Ruud Vinke; Craig S Levin
Journal:  Phys Med Biol       Date:  2015-06-17       Impact factor: 3.609

2.  Modelling the transport of optical photons in scintillation detectors for diagnostic and radiotherapy imaging.

Authors:  Emilie Roncali; Mohammad Amin Mosleh-Shirazi; Aldo Badano
Journal:  Phys Med Biol       Date:  2017-10-04       Impact factor: 3.609

3.  Predicting the timing properties of phosphor-coated scintillators using Monte Carlo light transport simulation.

Authors:  Emilie Roncali; Jeffrey P Schmall; Varsha Viswanath; Eric Berg; Simon R Cherry
Journal:  Phys Med Biol       Date:  2014-04-02       Impact factor: 3.609

4.  ScintSim1: A new Monte Carlo simulation code for transport of optical photons in 2D arrays of scintillation detectors.

Authors:  Mohammad Amin Mosleh-Shirazi; Zinat Zarrini-Monfared; Sareh Karbasi; Ali Zamani
Journal:  J Med Phys       Date:  2014-01
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.