Near real-time VOCs analysis using an aspiration ion mobility spectrometer.

Paweł Mochalski1, Joanna Rudnicka, Agapios Agapiou, Milt Statheropoulos, Anton Amann, Bogusław Buszewski.   

Abstract

The ChemPro 100i chemical detector (aspiration-type ion mobility spectrometer) was used for the detection of selected volatile organic compounds known to be potential indicators of human presence. The targeted group of compounds mainly comprised ketones (acetone, 2-butanone, 2-pentanone, 3-methyl-2-butanone, 4-heptanone), aldehydes (propanal, pentanal, hexanal, octanal), dimethyl disulfide (DMDS), isoprene and ethanol. Gaseous standards of these compounds were produced from pure substances and analysed using the aspiration ion mobility spectrometry (AIMS) chemical detector. The chemical fingerprints obtained (patterns) were compared to evaluate possible differences in responses. The limits of detection ranged from 5 ppbv for 4-heptanone to 87 ppbv for DMDS, whereas relative standard deviations varied from 1.5% to 5%. Additionally, quantitative AIMS measurements of acetone levels in human breath samples were carried out. The breath acetone levels measured with AIMS ranged from 290 to 540 ppbv and correlated quite well with the SPME-GC-MS results, showing limited potential of AIMS for the detection of breath acetone; however, deviations were observed for concentrations above 500 ppbv. The further success of AIMS in breath analysis depends on improvements of the analytical power (e.g. selectivity, sensitivity, resolution) and the implementation of multivariate data analysis techniques.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23470292     DOI: 10.1088/1752-7155/7/2/026002

Source DB:  PubMed          Journal:  J Breath Res        ISSN: 1752-7155            Impact factor:   3.262


  9 in total

Review 1.  Review on ion mobility spectrometry. Part 1: current instrumentation.

Authors:  R Cumeras; E Figueras; C E Davis; J I Baumbach; I Gràcia
Journal:  Analyst       Date:  2015-03-07       Impact factor: 4.616

2.  Assessment of the exhalation kinetics of volatile cancer biomarkers based on their physicochemical properties.

Authors:  Anton Amann; Pawel Mochalski; Vera Ruzsanyi; Yoav Y Broza; Hossam Haick
Journal:  J Breath Res       Date:  2014-02-24       Impact factor: 3.262

3.  Emission rates of selected volatile organic compounds from skin of healthy volunteers.

Authors:  Paweł Mochalski; Julian King; Karl Unterkofler; Hartmann Hinterhuber; Anton Amann
Journal:  J Chromatogr B Analyt Technol Biomed Life Sci       Date:  2014-04-13       Impact factor: 3.205

4.  Volatile organic compounds in ventilated critical care patients: a systematic evaluation of cofactors.

Authors:  Tobias Hüppe; Dominik Lorenz; Mario Wachowiak; Felix Maurer; Andreas Meiser; Heinrich Groesdonk; Tobias Fink; Daniel I Sessler; Sascha Kreuer
Journal:  BMC Pulm Med       Date:  2017-08-22       Impact factor: 3.317

5.  An In-Vitro Study for Early Detection and to Distinguish Breast and Lung Malignancies Using the Pcb Technology Based Nanodosimeter.

Authors:  P Venkatraman; C S Sureka
Journal:  Sci Rep       Date:  2019-01-23       Impact factor: 4.379

6.  Quantification of Volatile Acetone Oligomers Using Ion-Mobility Spectrometry.

Authors:  Tobias Hüppe; Dominik Lorenz; Felix Maurer; Tobias Fink; Ramona Klumpp; Sascha Kreuer
Journal:  J Anal Methods Chem       Date:  2021-08-02       Impact factor: 2.193

7.  Sensor Array and Gas Chromatographic Detection of the Blood Serum Volatolomic Signature of COVID-19.

Authors:  Yolande Ketchanji Mougang; Lorena Di Zazzo; Marilena Minieri; Rosamaria Capuano; Alexandro Catini; Jacopo Maria Legramente; Roberto Paolesse; Sergio Bernardini; Corrado Di Natale
Journal:  iScience       Date:  2021-07-10

Review 8.  Sensing Technologies for Detection of Acetone in Human Breath for Diabetes Diagnosis and Monitoring.

Authors:  Valentine Saasa; Thomas Malwela; Mervyn Beukes; Matlou Mokgotho; Chaun-Pu Liu; Bonex Mwakikunga
Journal:  Diagnostics (Basel)       Date:  2018-01-31

9.  Ion mobility spectrometry for pharmacokinetic studies--exemplary application.

Authors:  V Ruzsanyi
Journal:  J Breath Res       Date:  2013-11-28       Impact factor: 3.262

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.