Literature DB >> 23470208

Double cryoenergy application (freeze-thaw-freeze) at growing myocardium: lesion volume and effects on coronary arteries early after energy application. Implications for efficacy and safety in pediatric patients.

Heike E Schneider1, Maja Stahl, Thomas Kriebel, Wolfgang Schillinger, Manfred Schill, Johannes Jakobi, Thomas Paul.   

Abstract

INTRODUCTION: Data on radiofrequency current application (RFA) at growing myocardium suggest that coronary artery stenosis may occur with a low incidence. Cryoenergy has emerged as an effective alternative to RFA. Although already used in clinical practice, experience with cryoenergy is low due to lack of data concerning effects of a modified double cryoenergy application, a freeze-thaw-freeze cycle, at growing myocardium. Purpose of the study was to assess lesion volume (efficacy) and risk of coronary artery damage (safety) early, 48 hours, after modified double cryoenergy application in a piglet model. METHODS AND
RESULTS: In 5 piglets, following selective coronary angiography, two sequential cycles of cryoenergy were delivered at -75 °C for 4 minutes, interrupted by thawing for one minute, at the atrioventricular groove. Piglets were restudied after 48 hours by coronary angiography and intracoronary ultrasound (ICUS). Ablation lesions were examined morphologically and lesion volume was determined by 3-dimensional morphometric analysis. Lesion volume was 109.21 ± 39.61 mm(3) for atrial and 150.30 ± 53.21 mm(3) (P = 0.02) for ventricular lesions. Lesion depth was not significantly different for atrial, 3.07 ± 1.08 mm, versus ventricular lesions, 3.56 ± 1.3 mm. Cryoenergy induced minor coronary artery damage with medial and adventitial necrosis but a preserved intimal layer was present in 2/31 lesions, which had not been detected by coronary angiography or ICUS.
CONCLUSION: Early after double cryoenergy application, subclinical minor changes of the coronary artery wall could be detected occasionally whereas the intimal layer remained intact. These findings may have implications on efficacy and safety when cryoenergy is applied for tachycardia substrates in pediatric patients.
© 2013 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2013        PMID: 23470208     DOI: 10.1111/jce.12085

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  2 in total

1.  Comparison of 6-mm Versus 8-mm-Tip Cryoablation Catheter for the Treatment of Atrioventricular Nodal Reentrant Tachycardia in Children: A Prospective Study.

Authors:  Volkan Tuzcu; Enes Elvin Gul; Mehmet Karacan; Hacer Kamali; Nida Celik; Celal Akdeniz
Journal:  Pediatr Cardiol       Date:  2017-06-13       Impact factor: 1.655

2.  Effects of Triple Cryoenergy Application on Lesion Formation and Coronary Arteries in the Developing Myocardium.

Authors:  Ulrich Krause; Filipe D Abreu da Cunha; David Backhoff; Claudius Jacobshagen; Sophia Klehs; Heike E Schneider; Thomas Paul
Journal:  Pediatr Cardiol       Date:  2017-01-11       Impact factor: 1.655