Locking versus standard screw fixation for acetabular cups: is there a difference?

Snir Heller1, Tamar Brosh, Yona Kosashvili, Steven Velkes, Alon Burg, Israel Dudkiewicz.   

Abstract

INTRODUCTION: Total hip replacement has been established as a valid treatment option for displaced subcapital fractures. However, insufficient primary fixation may be the reason for early loosening in these osteoporotic patients. Primary fixation of the cup is usually achieved by press-fit fixation that can be enhanced using screws. Locking the screws into their respective cups may seem to improve the primary fixation of the construct, as locked plates proved superior fixation for osteoporotic fractures.
METHODS: The study consisted of three groups: in each group, three cups were fixed into blocks of foam bone using press-fit technique. In the first group, no additional screws were used, in the second group two standard screws were inserted, while in the third group two acetabular screws were cemented into the cup to simulate locked screw fixation. Load was applied onto the rim of the acetabular component to cause shearing between the cup and the block. Cup fixation was examined by a loading machine that acquired load versus displacement. The stiffness (load vs. displacement) was calculated.
RESULTS: Screws, either locked or non-locked, enhanced cup fixation by 26 % (p value <0.01). No significant changes were found between the locking and non-locking screws groups. DISCUSSION: These experimental results indicate that acetabular screws enhance primary cup fixation. This may become significant in conditions where the acetabular bone stock is suboptimal, such as when performing total hip arthroplasty after displaced subcapital fractures. However, there is no superiority for locked screws over standard screw fixation.

Entities:  

Mesh:

Year:  2013        PMID: 23467767     DOI: 10.1007/s00402-013-1709-6

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  6 in total

1.  Evaluation of the effect of custom burr holes on a surgeon's sense of screw fixation in revision porous metal cups.

Authors:  Mark A Nyland; Brent A Lanting; Hristo N Nikolov; Lyndsay E Somerville; Matthew G Teeter; James L Howard
Journal:  J Orthop       Date:  2015-10-19

2.  Minimum 2-Year Outcomes of a Novel 3D-printed Fully Porous Titanium Acetabular Shell in Revision Total Hip Arthroplasty.

Authors:  Elyse J Berlinberg; J Abraham Kavian; Mackenzie A Roof; Ittai Shichman; Brett Frykberg; William B Lutes; Erik A Schnaser; Stephen A Jones; Richard W McCalden; Ran Schwarzkopf
Journal:  Arthroplast Today       Date:  2022-10-12

Review 3.  The Diagnosis and Treatment of Acetabular Bone Loss in Revision Hip Arthroplasty: An International Consensus Symposium.

Authors:  Peter K Sculco; Timothy Wright; Michael-Alexander Malahias; Alexander Gu; Mathias Bostrom; Fares Haddad; Seth Jerabek; Michael Bolognesi; Thomas Fehring; Alejandro Gonzalez DellaValle; William Jiranek; William Walter; Wayne Paprosky; Donald Garbuz; Thomas Sculco
Journal:  HSS J       Date:  2021-09-28

4.  Are additional screws required for press-fit fixation of cementless acetabular cups? A systematic review and meta-analysis.

Authors:  Lei Guo; Tianlong Jiang; Shenghui Ni; Peng Luo
Journal:  J Orthop Traumatol       Date:  2022-02-10

5.  No effect of additional screw fixation of a cementless, all-polyethylene press-fit socket on migration, wear, and clinical outcome.

Authors:  Michiel J M Minten; Petra J C Heesterbeek; Maarten Spruit
Journal:  Acta Orthop       Date:  2016-06-14       Impact factor: 3.717

6.  Early aseptic loosening of the Tritanium primary acetabular component with screw fixation.

Authors:  William J Long; Samir Nayyar; Kevin K Chen; David Novikov; Roy I Davidovitch; Jonathan M Vigdorchik
Journal:  Arthroplast Today       Date:  2018-01-02
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.