Noise levels of neonatal high-flow nasal cannula devices--an in-vitro study.

Kai König1, Ellen L Stock, Melanie Jarvis.   

Abstract

BACKGROUND: Excessive ambient noise levels have been identified as a potential risk factor for adverse outcome in very preterm infants. Noise level measurements for continuous positive airway pressure (CPAP) devices demonstrated that these constantly exceed current recommendations. The use of high-flow nasal cannula (HFNC) as an alternative non-invasive ventilation modality has become more popular in recent years in neonatal care.
OBJECTIVE: To study noise levels of two HFNC devices commonly used in newborns. As a comparison, noise levels of a continuous flow CPAP device were also studied.
METHODS: In-vitro study. The noise levels of two contemporary HFNC devices (Fisher & Paykel NHF™ and Vapotherm Precision Flow®) and one CPAP device (Dräger Babylog® 8000 plus) were measured in the oral cavity of a newborn manikin in an incubator in a quiet environment. HFNC flows of 4-8 l/min and CPAP pressures of 4-8 cm H2O were applied. The CPAP flow was set at 8 l/min as per unit practice.
RESULTS: Vapotherm HFNC generated the highest noise levels, measuring 81.2-91.4 dB(A) with increasing flow. Fisher & Paykel HFNC noise levels were between 78.8 and 81.2 dB(A). The CPAP device generated the lowest noise levels between 73.9 and 77.4 dB(A).
CONCLUSIONS: Both HFNC devices generated higher noise levels than the CPAP device. All noise levels were far above current recommendations of the American Academy of Pediatrics. In light of the long duration of non-invasive respiratory support of very preterm infants, less noisy devices are required to prevent the potentially adverse effects of continuing excessive noise exposure in the neonatal intensive care unit.
Copyright © 2013 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2013        PMID: 23466653     DOI: 10.1159/000346764

Source DB:  PubMed          Journal:  Neonatology        ISSN: 1661-7800            Impact factor:   4.035


  4 in total

Review 1.  High-flow nasal cannula: recommendations for daily practice in pediatrics.

Authors:  Christophe Milési; Mathilde Boubal; Aurélien Jacquot; Julien Baleine; Sabine Durand; Marti Pons Odena; Gilles Cambonie
Journal:  Ann Intensive Care       Date:  2014-09-30       Impact factor: 6.925

2.  Investigating Noise Exposure to Newborn Infants From Respiratory Support: Methodological Considerations.

Authors:  Devika Singh; Gerhard Fusch
Journal:  Cureus       Date:  2021-11-08

Review 3.  Nasal high flow treatment in preterm infants.

Authors:  Calum T Roberts; Kate A Hodgson
Journal:  Matern Health Neonatol Perinatol       Date:  2017-09-06

4.  The comfort assessment in healthy adults during constant-flow mode in noninvasive ventilator.

Authors:  Juanjuan Yao; Wei Li; Mengmeng Peng; Kaixin He; Dedong Ma; Hongxiu Lu
Journal:  Clin Respir J       Date:  2021-12-06       Impact factor: 1.761

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.