An error analysis perspective for patient alignment systems.

Michael Figl1, Marcus Kaar, Rainer Hoffman, Alfred Kratochwil, Johann Hummel.   

Abstract

PURPOSE: This paper analyses the effects of error sources which can be found in patient alignment systems. As an example, an ultrasound (US) repositioning system and its transformation chain are assessed. The findings of this concept can also be applied to any navigation system. METHODS AND MATERIALS: In a first step, all error sources were identified and where applicable, corresponding target registration errors were computed. By applying error propagation calculations on these commonly used registration/calibration and tracking errors, we were able to analyse the components of the overall error. Furthermore, we defined a special situation where the whole registration chain reduces to the error caused by the tracking system. Additionally, we used a phantom to evaluate the errors arising from the image-to-image registration procedure, depending on the image metric used. We have also discussed how this analysis can be applied to other positioning systems such as Cone Beam CT-based systems or Brainlab's ExacTrac.
RESULTS: The estimates found by our error propagation analysis are in good agreement with the numbers found in the phantom study but significantly smaller than results from patient evaluations. We probably underestimated human influences such as the US scan head positioning by the operator and tissue deformation. Rotational errors of the tracking system can multiply these errors, depending on the relative position of tracker and probe.
CONCLUSIONS: We were able to analyse the components of the overall error of a typical patient positioning system. We consider this to be a contribution to the optimization of the positioning accuracy for computer guidance systems.

Entities:  

Mesh:

Year:  2013        PMID: 23463386     DOI: 10.1007/s11548-013-0819-5

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  25 in total

Review 1.  A review of calibration techniques for freehand 3-D ultrasound systems.

Authors:  Laurence Mercier; Thomas Langø; Frank Lindseth; D Louis Collins
Journal:  Ultrasound Med Biol       Date:  2005-04       Impact factor: 2.998

Review 2.  Advances in treatment techniques: arc-based and other intensity modulated therapies.

Authors:  Jian-Yue Jin; Ning Wen; Lei Ren; Carri Glide-Hurst; Indrin J Chetty
Journal:  Cancer J       Date:  2011 May-Jun       Impact factor: 3.360

3.  Clinical evaluation of a robotic 6-degree of freedom treatment couch for frameless radiosurgery.

Authors:  Thierry Gevaert; Dirk Verellen; Benedikt Engels; Tom Depuydt; Karina Heuninckx; Koen Tournel; Michael Duchateau; Truus Reynders; Mark De Ridder
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-09-22       Impact factor: 7.038

4.  A clinical comparison of patient setup and intra-fraction motion using frame-based radiosurgery versus a frameless image-guided radiosurgery system for intracranial lesions.

Authors:  Naren Ramakrishna; Florin Rosca; Scott Friesen; Evrim Tezcanli; Piotr Zygmanszki; Fred Hacker
Journal:  Radiother Oncol       Date:  2010-01-28       Impact factor: 6.280

5.  Automated dental implantation using image-guided robotics: registration results.

Authors:  Xiaoyan Sun; Frederic D McKenzie; Sebastian Bawab; Jiang Li; Yongki Yoon; Jen-K Huang
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-01-08       Impact factor: 2.924

6.  Quantification of organ motion during chemoradiotherapy of rectal cancer using cone-beam computed tomography.

Authors:  Irene Chong; Maria Hawkins; Vibeke Hansen; Karen Thomas; Helen McNair; Brian O'Neill; Alexandra Aitken; Diana Tait
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-06-12       Impact factor: 7.038

7.  Accuracy of ultrasound-based image guidance for daily positioning of the upper abdomen: an online comparison with cone beam CT.

Authors:  Judit Boda-Heggemann; Philipp Mennemeyer; Hansjörg Wertz; Nadja Riesenacker; Beate Küpper; Frank Lohr; Frederik Wenz
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-04-23       Impact factor: 7.038

8.  Prostate localization using transabdominal ultrasound imaging.

Authors:  Frieda Trichter; Ronald D Ennis
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-08-01       Impact factor: 7.038

9.  Image-guided radiotherapy for prostate cancer. Implementation of ultrasound-based prostate localization for the analysis of inter- and intrafraction organ motion.

Authors:  Michael Pinkawa; Martin Pursch-Lee; Branka Asadpour; Bernd Gagel; Marc D Piroth; Jens Klotz; Sandra Nussen; Michael J Eble
Journal:  Strahlenther Onkol       Date:  2008-12-24       Impact factor: 3.621

10.  Validation for 2D/3D registration. I: A new gold standard data set.

Authors:  S A Pawiro; P Markelj; F Pernus; C Gendrin; M Figl; C Weber; F Kainberger; I Nöbauer-Huhmann; H Bergmeister; M Stock; D Georg; H Bergmann; W Birkfellner
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

View more
  1 in total

1.  Evaluation of 3D ultrasound for image guidance.

Authors:  David Iommi; Johann Hummel; Michael Lutz Figl
Journal:  PLoS One       Date:  2020-03-26       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.