Comparison of cone beam computed tomography scans with and without simulation of head motion.

H Neller1, M A Geibel.   

Abstract

PURPOSE: To test for differences in image quality between 3D volumetric datasets acquired by cone beam computed tomography (CBCT) with and without simulation of head motion by visual analysis of individual image sections and by comparison of scan-based distance and Hounsfield unit (HU) measurements.
MATERIALS AND METHODS: A total of 200 volumetric datasets were acquired with and without remote-controlled, simulated movement of a human cadaver head using the KaVo 3D eXam CBCT system.
RESULTS: The "Landscape 8 x 8 cm Slide" mode provided a sufficient field of view at a low radiation dose. All datasets showed reproducible results. Our analysis showed that the level of image quality and image detail increased with increasing resolution. Linear distance measurements and HU measurements in the cone beam CT scans acquired without simulation of head motion were absolutely comparable to those obtained with head motion simulation.
CONCLUSIONS: There is no practical difference in image quality between CBCT scans acquired with and without head motion if no long-term change in head position occurs during the acquisition process. Slight head movement has no clinically relevant effect on the geometric accuracy or visual image quality of cone beam CT scans.

Entities:  

Mesh:

Year:  2012        PMID: 23457899

Source DB:  PubMed          Journal:  Int J Comput Dent        ISSN: 1463-4201            Impact factor:   1.883


  3 in total

1.  Movement characteristics in young patients and the impact on CBCT image quality.

Authors:  Rubens Spin-Neto; Louise H Matzen; Lars Schropp; Erik Gotfredsen; Ann Wenzel
Journal:  Dentomaxillofac Radiol       Date:  2016-02-26       Impact factor: 2.419

2.  An ex vivo study of automated motion artefact correction and the impact on cone beam CT image quality and interpretability.

Authors:  Rubens Spin-Neto; Louise H Matzen; Lars W Schropp; Thomas S Sørensen; Ann Wenzel
Journal:  Dentomaxillofac Radiol       Date:  2018-03-22       Impact factor: 2.419

3.  Patient movement characteristics and the impact on CBCT image quality and interpretability.

Authors:  Rubens Spin-Neto; Cláudio Costa; Daniela Mra Salgado; Nataly Rm Zambrana; Erik Gotfredsen; Ann Wenzel
Journal:  Dentomaxillofac Radiol       Date:  2017-10-20       Impact factor: 2.419

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.