Literature DB >> 23445712

Safety and effectiveness of an endoscopic suturing device in a human colonic treat-and-resect model.

Eric M Pauli1, Conor P Delaney, Bradley Champagne, Sharon Stein, Jeffrey M Marks.   

Abstract

BACKGROUND: Despite advances in many areas of therapeutic endoscopy, the development of an effective endoscopic suturing device has been elusive. The purpose of this trial was to evaluate the safety and effectiveness of a suturing device to place and secure sutures within normal, in vivo human colonic tissue prior to surgical resection.
METHODS: Patients undergoing elective colectomy were enrolled in this treat-and-resect model. The Overstitch endoscopic suturing device (Apollo Endosurgery, Austin, TX) was used to place sutures in healthy colonic tissue during a 15-minute, time-limited period. Following colectomy, the explanted tissue was evaluated to determine the depth of suture penetration and the effectiveness of the suture/cinch element. Clinical and operative data were recorded.
RESULTS: Four patients (50% female) were enrolled. Seven sutures were successfully placed, incorporating a total of 10 tissue bites in a mean of 13.5 minutes. On inspection of the explanted tissue, all sutures were found to be located subserosal (no full thickness bites were taken). The suture and cinch elements were judged to be effective in the majority of cases. One device-related issue did not inhibit the ability to oppose tissue or place the cinch. There were no intraoperative or postoperative complications.
CONCLUSIONS: The Overstitch permitted safe and effective suturing in an in vivo human colon model. The sutures were placed at a consistent subserosal depth and at no point risked iatrogenic injury to adjacent structures. Technical issues with the device were infrequent and did not inhibit the ability to place sutures effectively.

Entities:  

Keywords:  Overstitch; endolumenal surgery; endoscopic plication; endoscopic suturing

Mesh:

Year:  2013        PMID: 23445712     DOI: 10.1177/1553350613479204

Source DB:  PubMed          Journal:  Surg Innov        ISSN: 1553-3506            Impact factor:   2.058


  6 in total

Review 1.  Current applications of endoscopic suturing.

Authors:  Stavros N Stavropoulos; Rani Modayil; David Friedel
Journal:  World J Gastrointest Endosc       Date:  2015-07-10

Review 2.  Complications during colonoscopy: prevention, diagnosis, and management.

Authors:  R Manta; F Tremolaterra; A Arezzo; M Verra; G Galloro; L Dioscoridi; F Pugliese; A Zullo; M Mutignani; G Bassotti
Journal:  Tech Coloproctol       Date:  2015-07-11       Impact factor: 3.781

3.  The Efficacy of a Novel Tissue Grasper-Clips Technique for Large Perforations of the Sigmoid Colon in an Experimental Animal Model (Video).

Authors:  Jun Young Eun; Yunho Jung; Tae Hoon Lee; Young Sin Cho; Ho Sung Rhee; Young Kyu Jung; Joung-Ho Han; Duk Su Kim; Il Kwun Chung; Sang-Heum Park; Sun Joo Kim
Journal:  Dig Dis Sci       Date:  2017-02-08       Impact factor: 3.199

4.  Comprehensive management of full-thickness luminal defects: The next frontier of gastrointestinal endoscopy.

Authors:  Joshua S Winder; Eric M Pauli
Journal:  World J Gastrointest Endosc       Date:  2015-07-10

5.  Gastrointestinal exposed endoscopic full-thickness resection in the era of endoscopic suturing: a retrospective single-center case series.

Authors:  Antonino Granata; Alberto Martino; Michele Amata; Dario Ligresti; Mario Traina
Journal:  Wideochir Inne Tech Maloinwazyjne       Date:  2021-03-15       Impact factor: 1.195

Review 6.  Endoscopic Closure for Full-Thickness Gastrointestinal Defects: Available Applications and Emerging Innovations.

Authors:  Nobuyoshi Takeshita; Khek Yu Ho
Journal:  Clin Endosc       Date:  2016-08-30
  6 in total