Influence of distance between the rotation axis of back support and the hip joint on shear force applied to buttocks in a reclining wheelchair's back support.

Kenichi Kobara1, Daisuke Fujita, Hiroshi Osaka, Tomotaka Ito, Susumu Watanabe.   

Abstract

PURPOSE: The purpose of this study was to investigate the influence of the distance between the position of the rotation axis of the back support and hip joint on the changes in the shear force applied to the buttocks. STUDY
DESIGN: Repeated-measures design.
METHODS: The subjects were 11 healthy adult men without leg and/or trunk diseases. The shear force applied to the buttocks was measured using a force plate. This study had three experimental conditions: the position of hip joint was taken as the 3 cm, 6 cm, and 9 cm forward from the standard sitting position.
RESULTS: In returning to an upright position of back support, the shear force was 15.0 ± 2.9 %BW in the 3 cm forward condition, 16.7 ± 3.6 %BW in the 6 cm forward condition, and 19.5 ± 5.3 %BW in the 9 cm forward condition. Significant differences appeared between the 9 cm and the other conditions (p < 0.05).
CONCLUSIONS: The results of this study suggested that the shear force applied to the buttocks changed in reclining back support and an increase in the distance between the axis of rotation of the back support and hip joint led to an increase in the remaining shear force after reclining the back support.

Entities:  

Keywords:  Decubitus ulcer; shear force

Mesh:

Year:  2013        PMID: 23436694     DOI: 10.1177/0309364613476534

Source DB:  PubMed          Journal:  Prosthet Orthot Int        ISSN: 0309-3646            Impact factor:   1.895


  3 in total

1.  Risk factors associated with intraoperatively acquired pressure ulcers in the park-bench position: a retrospective study.

Authors:  Mine Yoshimura; Shinji Iizaka; Michihiro Kohno; Osamu Nagata; Takashi Yamasaki; Tomoko Mae; Naoko Haruyama; Hiromi Sanada
Journal:  Int Wound J       Date:  2015-06-04       Impact factor: 3.315

2.  Effect of asymmetrical buttock pressure on forces exerted in buttock movement when using a reclining chair.

Authors:  Yasuyuki Nagata; Kenichi Kobara; Daisuke Fujita
Journal:  J Phys Ther Sci       Date:  2021-05-15

3.  Effect of rotational axis position of wheelchair back support on shear force when reclining.

Authors:  Kenichi Kobara; Hiroshi Osaka; Hisashi Takahashi; Tomotaka Ito; Daisuke Fujita; Susumu Watanabe
Journal:  J Phys Ther Sci       Date:  2014-05-29
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.