Literature DB >> 23427578

[Optical design and performance analysis of light and small echelle spectrograph].

Chun-Li Ning1, Xiang-Dong Qi, Shao-Jie Chen, Ji-Cheng Cui.   

Abstract

The present paper analyzed the influence of the pinhole diameter, grating parameters, CCD pixel size, prism parameters, system aberrations and found that the first three are the main factors, and then deduced the mutual restraint relationship among them. On this basis, a portable high-resolution echelle spectrograph was designed. Applying this design, aberration was fully corrected and spectral resolution achieved the demand. With the Hg lamp calibrated and restored, the actual resolution gets up to 0.038 nm which is significantly better than target (0.05@200 nm), while the ordinary grating spectrometers need 500 mm focal length to achieve this resolution. From this result, the advantages of the portable echelle spectrograph are fully demonstrated.

Entities:  

Year:  2012        PMID: 23427578

Source DB:  PubMed          Journal:  Guang Pu Xue Yu Guang Pu Fen Xi        ISSN: 1000-0593            Impact factor:   0.589


  1 in total

1.  Design of Compact Mid-Infrared Cooled Echelle Spectrometer Based on Toroidal Uniform-Line-Spaced (TULS) Grating.

Authors:  Qingyu Wang; Honghai Shen; Weiqi Liu; Jingzhong Zhang; Lingtong Meng
Journal:  Sensors (Basel)       Date:  2022-09-26       Impact factor: 3.847

  1 in total