Literature DB >> 23427212

Evaluation of a novel closed-loop total intravenous anaesthesia drug delivery system: a randomized controlled trial.

T M Hemmerling1, E Arbeid, M Wehbe, S Cyr, R Taddei, C Zaouter.   

Abstract

BACKGROUND: We have developed an automatic anaesthesia system for closed-loop administration of anaesthesia drugs. The control variables used were bispectral index (BIS) and Analgoscore for hypnosis and antinociception, respectively.
METHODS: One hundred and eighty-six patients were randomly enrolled in two groups. Propofol, remifentanil, and rocuronium were administered using closed-loop feedback control (closed-loop, n = 93) or manually (control group, n = 93). The clinical performance of hypnosis control was determined by calculating the offset from a BIS of 45: 'excellent', 'good', 'poor', and 'inadequate' control was defined as BIS values within 10%, from 11% to 20%, from 21% to 30%, or >30% offset from the target. The clinical performance of analgesia was defined as the offset from Analgoscore values. Data presented as mean (standard deviation) (95% confidence interval).
RESULTS: Excellent or good control of hypnosis was achieved significantly longer in the closed-loop group [47.0 (9.8%) (45.0/49.0), 34.4 (4.7%) (33.5/35.4)] than in the control group [37.3 (14.3%) (34.4/40.2) and 32.3 (7.6%) (30.7/33.7)], respectively (P<0.0001 and 0.0085). Poor and inadequate control of hypnosis was significantly shorter in the closed-loop group [10.8 (5.0%) (9.8/11.8) and 7.7 (6.2%) (6.4/9.0)] than in the control group [14.7 (6.8%) (13.3/16.0) and 15.8 (14.7%) (12.8/18.8)], respectively (P<0.0001). Excellent control of analgesia was achieved significantly longer in the closed-loop group [78.7 (16.2%) (75.4/82.0)] than in the control group [73.7 (17.8%) (70.1/77.3)] (P=0.0456).
CONCLUSIONS: The closed-loop system was better at maintaining BIS and Analgoscore than manual administration.

Entities:  

Keywords:  McSleepy; closed-loop systems

Mesh:

Substances:

Year:  2013        PMID: 23427212     DOI: 10.1093/bja/aet001

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  14 in total

1.  Design of multichannel functional near-infrared spectroscopy system with application to propofol and sevoflurane anesthesia monitoring.

Authors:  Zhenhu Liang; Yue Gu; Xuejing Duan; Lei Cheng; Shujuan Liang; Yunjie Tong; Xiaoli Li
Journal:  Neurophotonics       Date:  2016-10-05       Impact factor: 3.593

Review 2.  Automation of anaesthesia: a review on multivariable control.

Authors:  Jing Jing Chang; S Syafiie; Raja Kamil; Thiam Aun Lim
Journal:  J Clin Monit Comput       Date:  2014-06-25       Impact factor: 2.502

3.  A novel system for automated propofol sedation: hybrid sedation system (HSS).

Authors:  Cedrick Zaouter; Riccardo Taddei; Mohamad Wehbe; Erik Arbeid; Shantale Cyr; Francesco Giunta; Thomas M Hemmerling
Journal:  J Clin Monit Comput       Date:  2016-03-12       Impact factor: 2.502

Review 4.  Intelligent automated drug administration and therapy: future of healthcare.

Authors:  Richa Sharma; Dhirendra Singh; Prerna Gaur; Deepak Joshi
Journal:  Drug Deliv Transl Res       Date:  2021-01-14       Impact factor: 4.617

5.  Closed-loop control better than open-loop control of profofol TCI guided by BIS: a randomized, controlled, multicenter clinical trial to evaluate the CONCERT-CL closed-loop system.

Authors:  Yu Liu; Min Li; Dong Yang; Xuena Zhang; Anshi Wu; Shanglong Yao; Zhanggang Xue; Yun Yue
Journal:  PLoS One       Date:  2015-04-17       Impact factor: 3.240

Review 6.  Newer drug delivery systems in anesthesia.

Authors:  Sona Dave; Deepa Shriyan; Pinakin Gujjar
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2017 Apr-Jun

Review 7.  Anaesthesia in austere environments: literature review and considerations for future space exploration missions.

Authors:  Matthieu Komorowski; Sarah Fleming; Mala Mawkin; Jochen Hinkelbein
Journal:  NPJ Microgravity       Date:  2018-02-23       Impact factor: 4.415

8.  Technical and clinical evaluation of a closed loop TIVA system with SEDLineTM spectral density monitoring: Multicentric prospective cohort study.

Authors:  Claudia Castellanos Peñaranda; Fabián D Casas Arroyave; Francisco J Gómez; Paola A Pinzón Corredor; Juan M Fernández; Marcela Velez Botero; Juan D Bohórquez Bedoya; Carlos Marulanda Toro
Journal:  Perioper Med (Lond)       Date:  2020-01-09

9.  Performance measurement of intraoperative systolic arterial pressure to predict in-hospital mortality in adult liver transplantation.

Authors:  Hyung-Chul Lee; Ho-Geol Ryu; Chul-Woo Jung
Journal:  Sci Rep       Date:  2017-08-01       Impact factor: 4.379

Review 10.  Recent advances in intravenous anesthesia and anesthetics.

Authors:  Mohamed Mahmoud; Keira P Mason
Journal:  F1000Res       Date:  2018-04-17
View more