We propose and experimentally demonstrate an ultra-thin rigid endoscope (450 μm diameter) based on a passive multimode optical fiber. We use digital phase conjugation to overcome the modal scrambling of the fiber to tightly focus and scan the laser light at its distal end. By exploiting the maximum number of modes available, sub-micron resolution, high quality fluorescence images of neuronal cells were acquired. The imaging system is evaluated in terms of fluorescence collection efficiency, resolution and field of view. The small diameter of the proposed endoscope, along with its high quality images offer an opportunity for minimally invasive medical endoscopic imaging and diagnosis based on cellular phenotype via direct tissue penetration.
We propose and experimentally demonstrate an ultra-thin rigid endoscope (450 μm diameter) based on a passive multimode optical fiber. We use digital phase conjugation to overcome the modal scrambling of the fiber to tightly focus and scan the laser light at its distal end. By exploiting the maximum number of modes available, sub-micron resolution, high quality fluorescence images of neuronal cells were acquired. The imaging system is evaluated in terms of fluorescence collection efficiency, resolution and field of view. The small diameter of the proposed endoscope, along with its high quality images offer an opportunity for minimally invasive medical endoscopic imaging and diagnosis based on cellular phenotype via direct tissue penetration.
Entities:
Keywords:
(070.5040) Phase conjugation; (090.1995) Digital holography; (110.2350) Fiber optics imaging; (170.0110) Imaging systems; (170.2150) Endoscopic imaging; (170.7050) Turbid media
Endoscopy has revolutionized biomedical imaging. The ability to directly visualize structures of
living organisms deep inside the body has proven indispensable both for prognosis and diagnosis. In
the state of the art of endoscopic imaging one can identify two distinct approaches [1, 2]. The first
relies on the miniaturization of optical elements like Gradient Index (GRIN) lenses [3-6], whereas
on the other hand there is a different approach that utilizes fiber optics to achieve the goal of
imaging deep inside the tissue. In the fiber optics domain, a large number of commercially available
endoscopes are built using fiber bundles, where each fiber acts as a single pixel of the captured
image [7, 8]. On the other hand there has been research on hybrid assemblies that exploit
single-mode and multimode optical fibers, mechanical actuators and lens systems to deliver and scan
a focused spot and collect the fluorescent light. This approach delivers highly versatile flexible
endoscopes that can perform one and two photon imaging [9-11]. The major disadvantage is that
the smallest achieved diameter is still a few millimeters.In an attempt to shrink the size of fiber-based endoscopes, multimode fibers with their large
number of degrees of freedom (directly linked to the number of propagation modes supported) appear
as very suitable candidates. However, multimode fibers being the waveguide equivalent of free space
scattering media, tend to scramble any information that is transmitted through them both in space
and time [12]. The optical field that enters
the fiber gets coupled to the different propagation modes, which follow different paths along the
propagation, possibly exchanging energy and reach the output, were the field appears as a random
pattern. Different techniques have been proposed to undo the scrambling properties of multimode
fibers and to focus and scan light through them, like direct-search wavefront shaping techniques
[13] or other complex optimization algorithms
[14] and Digital Phase Conjugation
[15]. Recently, different groups have
presented fluorescent and linear scattering imaging based on multimode fibers [16-18].In the present paper, we demonstrate a sub-micron resolution, lensless scanning fluorescent
imaging scheme by using a multimode fiber and digital phase conjugation to undo the scattering
properties of the waveguide. In our approach, we exploit all the available modes in a very high
numerical aperture (NA) multimode fiber, in order to generate and scan, optimal resolution and
contrast, focused spots and eventually to capture high-detail images of stained biological samples.
Optical phase conjugation is a powerful technique that has been proposed and implemented for
compensating the scattering that turbid media induce on imaging [19-23] along with
compensating the modal and polarization dispersion in multimode fibers [24-26]. Digital phase
conjugation [27], substitutes the nonlinear
crystal that is needed for conventional phase conjugation with the pair of a digital sensor and a
Spatial Light Modulator (SLM). Compared to iterative based wavefront optimization techniques,
digital phase conjugation (DPC) requires only a single shot measurement. DPC extracts the correct
phase front that undoes the modal scrambling created by the multimode fiber and generates a tightly
focused spot at the distal tip of the fiber. The system is open loop; therefore the time that is
needed for the calibration of the system is minimal.
2. Principle of operation and experimental setup
The proposed endoscopic system works as a reflection scanning fluorescence microscope
[28]. In order to obtain an image, a focused
spot is generated at the distal tip of the fiber and scanned on the object. The induced fluorescence
from each point of the scanning grid is collected by the same optical system and is integrated on a
photodetector. In our approach, we exploit the large number of degrees of freedom intrinsically
available in the multimode fiber (directly linked to number of propagation modes supported by the
fiber waveguide) to perform all the described actions. Initially, light is focused on the distal tip
of the fiber and gets coupled to the different propagation modes. These modes reach the output of
the fiber after following different paths inside the fiber core and also exchanging energy between
each other, generating finally what is seemingly a random speckle pattern. The speckled output is
combined with the reference and the resulting interference pattern is digitally recorded on the CMOS
sensor. We can then reconstruct the field at the proximal end of the fiber and calculate the
corresponding conjugate phase pattern. The reference beam transmitted through the beamsplitter is
incident onto the SLM, and gets modulated by the phase conjugate pattern displayed on the SLM. The
optical conjugate field that is formed in this way is reflected again on the beamsplitter and
follows the same path as the image, in the reverse direction, retracing its way back through the
fiber to exit the waveguide generating finally a tightly focused diffraction-limited spot at the
original position.A phase lookup table is generated for all the different points along a regular orthogonal grid by
scanning the calibration beam around the fiber facet and reconstructing the phase pattern for each
corresponding spot (as described above). The generation of diffraction limited spots, along the
regular grid allows us to perform scanning based imaging. Light that is tightly focused onto the
sample induces a fluorescent signal from the excited volume. The signal is emitted isotropically (in
an ideal model where the excited volume is reasonably small) and part of the generated signal is
captured by the fiber. The fluorescent signal that is captured by the multimode fiber propagates
backwards and is collected by the same CMOS sensor that was used for the imaging, which now acts as
a photon bucket. In the case when the photon budget is low, the signal from the fiber can be
diverted towards a Photomultiplying Tube (PMT) or any other sensitive point photodetector. The
experimental setup is described and graphically presented on Fig.
1.
Fig. 1
Experimental setup and picture of the needle endoscope. On the lower part, the experimental setup
is presented. Light focused on the fiber tip is dispersed over the different propagation modes
supported by the fiber (inset). The speckled output is combined with the reference and the resulting
interference pattern is digitally recorded onto the CMOS sensor. The reconstructed phase of the
hologram is assigned on the Spatial Light Modulator, which then modulates the reference beam. The
phase conjugate beam that is generated propagates backwards, retracing its way through the multimode
fiber and coming into focus at the original position. The calibration beam can be scanned in both
lateral dimensions (moving the calibration objective using motorized stages) generating a regular
grid of focused spots and the corresponding phase lookup table. Upper part. Placing the fiber inside
a needle tip we can make a rigid ultrathin endoscope that will allow the fiber to remain intact and
at the same time allow for minimally invasive endoscopic imaging. The outer diameter of the
endoscope is limited only by the needle and can be as small as 460μm for
the 250μm cladding fiber used in the experiments.
Experimental setup and picture of the needle endoscope. On the lower part, the experimental setup
is presented. Light focused on the fiber tip is dispersed over the different propagation modes
supported by the fiber (inset). The speckled output is combined with the reference and the resulting
interference pattern is digitally recorded onto the CMOS sensor. The reconstructed phase of the
hologram is assigned on the Spatial Light Modulator, which then modulates the reference beam. The
phase conjugate beam that is generated propagates backwards, retracing its way through the multimode
fiber and coming into focus at the original position. The calibration beam can be scanned in both
lateral dimensions (moving the calibration objective using motorized stages) generating a regular
grid of focused spots and the corresponding phase lookup table. Upper part. Placing the fiber inside
a needle tip we can make a rigid ultrathin endoscope that will allow the fiber to remain intact and
at the same time allow for minimally invasive endoscopic imaging. The outer diameter of the
endoscope is limited only by the needle and can be as small as 460μm for
the 250μm cladding fiber used in the experiments.We define the working distance of the MMF endoscope as the distance of the focusing plane away
from the distal tip of the multimode fiber, and the Field of View (FOV) as the maximum x and y
dimensions of the corresponding regular grid along which the focus can be scanned. The distance
between two adjacent focused spots is chosen based on the desired resolution and the speed of the
system. The resolution is limited by the diffraction limit imposed by the fiber numerical aperture.
Following a geometrical optics analysis (see Appendix A), we
can estimate the effective numerical aperture of the fiber for each point in front of the fiber
facet, as a function of the working distance (distance from the fiber facet, z) and distance from
the optical axis, x. The resolution and fluorescence collection efficiency of the optical system as
a function of the NA(x,z) at each point are given by, and the fluorescence collection efficiency, The simulation results for both the resolution and the fluorescence
collection efficiency are presented in Fig. 2.
Fig. 2
Resolution and fluorescence collection efficiency of the fiber endoscope as a function of working
distance z and the distance from the optical axis of the system x as calculated using a geometrical
optics analysis (Appendix A). The calculations are done for a
220μm diameter, 0.53 NA fiber. As shown in Fig. 2(a) and (c), the resolving power of the imaging system becomes worse as the imaging
plane is set away from the fiber tip while at the same time, the useful field of view increases.
indicate a clear trade off between resolution and field of view. Fig
2 (b), (d) present the fluorescence collection efficiency of the imaging system as a function
of position. The efficiency is worse as the fluorophore is placed away from the fiber facet and
towards the edges of the field of view. The simulations can help us predict the behavior of the
system and correct any differences in the fluorescent levels of the final image.
Resolution and fluorescence collection efficiency of the fiber endoscope as a function of working
distance z and the distance from the optical axis of the system x as calculated using a geometrical
optics analysis (Appendix A). The calculations are done for a
220μm diameter, 0.53 NA fiber. As shown in Fig. 2(a) and (c), the resolving power of the imaging system becomes worse as the imaging
plane is set away from the fiber tip while at the same time, the useful field of view increases.
indicate a clear trade off between resolution and field of view. Fig
2 (b), (d) present the fluorescence collection efficiency of the imaging system as a function
of position. The efficiency is worse as the fluorophore is placed away from the fiber facet and
towards the edges of the field of view. The simulations can help us predict the behavior of the
system and correct any differences in the fluorescent levels of the final image.Figures 2(a) and (c) demonstrate the resolving efficiency
of the system, and demonstrate the tradeoff between resolution and field of view as a function of
the working distance. The resolution degrades as we move away from the fiber tip but the field of
view that we can image becomes larger. Figures 2 (b) and (d)
show the relation between fluorescence collection efficiency and the imaging position and follow the
same behavior as the resolution. The collection efficiency worsens as we image farther away from the
fiber facet and towards the edges of the field of view.
3. Results and discussion
The performance of the fiber-based endoscope was primarily evaluated for a field of view that is
larger compared to any of the reported imaging techniques using multimode fibers [16, 17]. In Fig. 3 we present the results of imaging a fluorescent pattern on a
glass slide prepared with photolithography. The lateral size of the object is
100μm × 40μm and the total field of view
in both images is 110μm × 100μm. The image
on the left is the direct widefield fluorescent image recorded in reflection on the camera CMOS2
(see the experimental setup in Fig. 1) whereas the image on
the right corresponds to the scanning fluorescence image of the same sample captured through the
fiber where the pixel pitch was set to 1μm. The working distance of the
endoscope was set to 200μm. We observe that the image quality is very good
across the whole field of view without any apparent aberrations introduced. However, a difference on
the level of the fluorescence signal between the centre and the edges of the FOV can be observed and
matches well with the predicted fluorescence collection efficiency calculations presented in Fig. 2.
Fig. 3
Fiber endoscope large field of view imaging. The scanning fluorescence image captured from a
large field of view with the fiber endoscope at a working distance of 200m is a high fidelity
replica of the widefield image taken with a microscope objective. The difference in the fluorescence
level between the center and the edges is attributed to the different collections efficiencies away
from the optical axis of the fiber and the quasi linear dependence observed coincides with the
predicted behavior for the specific working distance (see Fig.
2(d)). The endoscope is capable of providing high information capacity, large field of view
images. Scale bar equal to 20μm.
Fiber endoscope large field of view imaging. The scanning fluorescence image captured from a
large field of view with the fiber endoscope at a working distance of 200m is a high fidelity
replica of the widefield image taken with a microscope objective. The difference in the fluorescence
level between the center and the edges is attributed to the different collections efficiencies away
from the optical axis of the fiber and the quasi linear dependence observed coincides with the
predicted behavior for the specific working distance (see Fig.
2(d)). The endoscope is capable of providing high information capacity, large field of view
images. Scale bar equal to 20μm.Next, we assess the image quality of the proposed system, first by comparing it against the image
quality produced by a lens based optical system of a comparable numerical aperture and then by
evaluating the resolution limit. For this purpose, we use a sample that consists of
1μm fluorescent beads laid on a glass substrate. This type of sample will
introduce a higher complexity and richer information content. In Fig. 3(a) and (d), we compare the image of the fluorescent beads as taken with a
20×, 0.5 NA microscope objective (similar to the NA of the multimode fiber used in the
experiments) against the image acquired with the scanning fluorescense endoscope. In order to
directly compare the two images, the image acquired with the scanning method is resampled and
convolved with a gaussian filter to eliminate the pixelation effect. The field of view is
60μm × 60μm in both images and the pixel
pitch for the fiber scanning is set to 1μm. The level of detail as conveyed
by both images is the same, as all possible features that are visible on the microscope objcetive
image can be identified on the scanning image taken with the fiber. The overall differences on the
level of the fluorescent signal are attributed to the different sides that the images are taken.
Moreover, we present cross-sectional plots from two different parts of the image so that we can have
a more quantitative comparison. We can observe in Fig.
3(b),(e) and Fig. 3(c),(f) that there is an one to one
correspondence of the images acquired. As a conclusion, the lensless, fiber-based scanning
fluorescence endoscope that we demonstrate can provide imaging quality that is directly comparable
to a lens based optical system of the same numerical aperture.In order to quantify the resolution limit of the fiber endoscope, we take a more detailed look
into the bead sample as shown in Fig. 4(g, h). In this case
the widefield fluorescence image is taken with a 0.65 NA microscope objective so that a more
detailed view of the 1μm bead sample is revealed. For the scanning system, the pixel pitch is set to
0.5μm, smaller than the expected resolution and the field of view is set to
30μm by 30μm. We can observe that beads closely
placed to each other at a distance smaller than 1μm (lower right part of the highlighted box) can be
still resolved with our endoscopic system therefore setting the resolution limit of the system to
the submicron range, significantly smaller than anything reported thus far [16-18].
Fig. 4
Comparison of the fiber endoscopic imaging modality vs a conventional optical system of the same
numerical aperture and estimation of resolution limit. (a) Widefield 0.5NA microscope objective
image and (d) Image acquired with the multimode fiber. Comparison of the imaging capability of the
system against a conventional optical system of similar NA. The information conveyed by the lensless
multimode fiber endoscope is directly comparable to that given by the 0.5 NA microscope objective.
The comparison of the cross sectional plots (b),(e) and (c),(f) verifies the conclusion. Fig. 3(g) and (h) are used to estimate the resolution of the
system. In Fig. 3(g) the bead sample is imaged with a 0.65 NA
objective to get a more detailed view. The scanning scheme (with a pixel pitch of
0.5μm) gives an image where two beads almost fused into each other can be
still be separated, therefore placing the resolution limit of the system in the submicron range.
Scale bars in all images are 10μm.
Comparison of the fiber endoscopic imaging modality vs a conventional optical system of the same
numerical aperture and estimation of resolution limit. (a) Widefield 0.5NA microscope objective
image and (d) Image acquired with the multimode fiber. Comparison of the imaging capability of the
system against a conventional optical system of similar NA. The information conveyed by the lensless
multimode fiber endoscope is directly comparable to that given by the 0.5 NA microscope objective.
The comparison of the cross sectional plots (b),(e) and (c),(f) verifies the conclusion. Fig. 3(g) and (h) are used to estimate the resolution of the
system. In Fig. 3(g) the bead sample is imaged with a 0.65 NA
objective to get a more detailed view. The scanning scheme (with a pixel pitch of
0.5μm) gives an image where two beads almost fused into each other can be
still be separated, therefore placing the resolution limit of the system in the submicron range.
Scale bars in all images are 10μm.Having verified the efficacy of the scanning fluorescent fiber imaging, we demonstrate high
resolution imaging of stained biological samples. This type of samples can be considered, as the
real standard that will define whether the system can be used as an endoscope for in vivo diagnosis
based on cellular phenotype. For this purpose, we prepare a sample of fluorescently labeled neuronal
cells.Probing cellular and subcellular details with the multimode fiber endoscope. Images of
fluorescently stained neuronal cells acquired with the multimode fiber endoscope and compared
against conventional images acquired with a microscope objective. First column, (a) widefield
fluorescent image of a single neuron soma and (d) detail of dendrites, middle row; (b) and (e)
direct stitched image as acquired from the fiber and right row, image from the fiber resampled and
filtered so that the pixelation induced by the scanning acquisition is overcome. Highly detailed
images of the neuronal soma and the dendritic network can be resolved by the fiber imaging system.
The high quality of the images can make this endoscope useful for diagnostic purposes based on
cellular phenotype. The working distance is 200μm to compensate for the
coverslip that separates the cells from the fiber facet. Field of view is
60μm by 60μm and scale bars in all images are
10μm.In Fig. 6, we present the images of two different parts of
the sample. The left column consists of the images as taken with the CMOS2 camera in reflection
under widefield excitation (40×, 0.65NA objective), the middle column are the images as they
are assembled by digitally scanning the focus spot, pixel by pixel on the regular grid and finally
in the right-hand column are the scanned images after resampling and Gaussian filtering to overcome
the pixelation that the scanning induces. The acquired images showing the neuronal soma in Fig. 6(b, c) carry all the structural information of the cell body
as can be seen by the widefield image Fig. 6(a). The quality
of the acquired images can lead to effective diagnosis that depends on cellular phenotype. On the
lower row, we present images of more detailed structures; dendrites that have a diameter of
1–3μm. In this case, where the photon budget is lower, the image
quality starts degrading; however, we can still collect interesting information about the
conformation of such small features.
Fig. 6
Geometrical optics calculations of the beam diameter as a function of the distance x from the
optical axis and the working distance z.
Geometrical optics calculations of the beam diameter as a function of the distance x from the
optical axis and the working distance z.
4. Conclusions and outlook
We have demonstrated that by exploiting all the degrees of freedom available in a multimode fiber
using digital phase conjugation to focus and scan the light, we have achieved sub-micrometer
resolution scanning fluorescent imaging of biological samples. The resolution of the system has been
measured by distinguishing 1μm fluorescent beads and was identified to
reside on the sub–micron range. The system has been used to image fluorescently stained
neuronal cells and detailed images successfully revealed the neuronal somata and the dendritic mesh.
The quality of the images offers an opportunity for direct diagnosis based on cellular phenotype.
However, the major limitation of multimode fibers is the modal distribution dependence on the
spatial conformation (bending) of the fiber. This problem is present in all imaging systems that
rely on multimode fibers both for the delivery and the collection of light, as the calibrated phase
look-up table will generate correct results only for small perturbations of the fiber spatial
conformation [18]. One possible solution to
overcoming the problems associated with bending is to make the fiber rigid (therefore avoiding the
bending) by epoxying it inside a needle tip. The multimode fiber was inserted in a 25-gauge needle
(450 μm diameter), thereby demonstrating a significant reduction in size
compared to other endoscopes. Moreover, the fiber part of the endoscope is passive (no active parts)
which is expected to facilitate its acceptance since most of fiber endoscopes used clinically are
based on passive fiber bundles. An endoscope with a rigid needle front-end is suitable for many
applications and the device described in this paper suits this need, if care is taken to provide
mechanical stability to the entire fiber. In such applications the proposed fiber endoscope might
open opportunities for minimally invasive high-resolution endoscopic imaging through direct tissue
penetration. For applications where a flexible probe is required, then we need to find ways to
compensate for the spatial mode scrambling. We are working on two separate solutions to this
problem. The first approach is to parameterize the spatial configuration of the fiber and accumulate
enough input-output data to be able to dynamically calibrate the instrument. The second approach is
based on the use of a point source (e.g. a metal nanoparticle) at the distal tip of the fiber, which
allows us to continuously recalculate the required excitation light distribution in order to
maintain focus.