Fiberscopic estimation of vocal fold stiffness in vivo using the sucking method.

S Tanaka1, M Hirano.   

Abstract

We developed a fiberscopic device for estimating vocal fold stiffness in vivo. A small plastic tube with a hole in the side was connected to the forceps channel of the fiberscope and used as an aspiration tube under constant negative pressure. A basic experiment revealed that the maximum distance at which the vocal fold was sucked to the hole depended on the stiffness of the tissue. A preliminary study was conducted with a normal subject and four patients: one with sulcus vocalis, one with Reinke's edema, and two with carcinoma. The stiffness estimated using this method was 3 to 4 g/cm for the normal vocal fold. The stiffness for carcinoma and sulcus vocalis was significantly greater than that of normal vocal fold tissue. The stiffness of Reinke's edema did not differ markedly from that of normal tissue.

Entities:  

Mesh:

Year:  1990        PMID: 2340125     DOI: 10.1001/archotol.1990.01870060079015

Source DB:  PubMed          Journal:  Arch Otolaryngol Head Neck Surg        ISSN: 0886-4470


  3 in total

1.  Characterization of lamina propria and vocal muscle in human vocal fold tissue by ultrasound Nakagami imaging.

Authors:  Po-Hsiang Tsui; Chih-Chung Huang; Lei Sun; Seth H Dailey; K Kirk Shung
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

Review 2.  Functional assessment of the ex vivo vocal folds through biomechanical testing: A review.

Authors:  Gregory R Dion; Seema Jeswani; Scott Roof; Mark Fritz; Paulo G Coelho; Michael Sobieraj; Milan R Amin; Ryan C Branski
Journal:  Mater Sci Eng C Mater Biol Appl       Date:  2016-04-08       Impact factor: 7.328

3.  In vivo measurement of the elastic properties of the human vocal fold.

Authors:  Eric Goodyer; Frank Muller; Brian Bramer; Dilip Chauhan; Markus Hess
Journal:  Eur Arch Otorhinolaryngol       Date:  2006-02-22       Impact factor: 2.503

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.