Evaluation of an electromagnetic image-fusion navigation system for biopsy of small lesions: assessment of accuracy in an in vivo swine model.

Liat Appelbaum1, Luigi Solbiati, Jacob Sosna, Yzhak Nissenbaum, Nathalie Greenbaum, S Nahum Goldberg.   

Abstract

PURPOSE: To evaluate the accuracy of a novel combined electromagnetic (EM) navigation/image fusion system for biopsy of small lesions.
MATERIALS AND METHODS: Using ultrasound (US) guidance, metallic (2 × 1 mm) targets were imbedded in the paraspinal muscle (n = 28), kidney (n = 18), and liver (n = 4) of four 55- to 65-kg pigs. Baseline helical computed tomography (CT) imaging (Brilliance; Philips) identified these biopsy targets and six and nine cutaneous fiducial markers. CT data were imported into a MyLab Twice system (Esaote, Genoa, Italy) for CT/US image fusion. After verification of successful image fusion, baseline registration error and respiratory motion error were assessed by documenting deviation of the US and CT position of the targets in real time. Biopsy targeting was subsequently performed under conditions of normal respiratory using 15-cm 16G eTrax needles (Civco). To mimic the conditions of poor US visualization, only reconstructed CT information was displayed during biopsy. Accuracy of targeting was measured by repeat CT scanning as the distance of the needle tip to the target center. Targeting accuracy of free-hand vs. guided technique, and electromagnetic (EM) sensor positioning (ie, on the hub or within the needle stylus tip) were evaluated.
RESULTS: In muscle, needle registration error was 0.9 ± 1.2 mm and respiratory motion error 4.0 ± 1.0 mm. Target accuracy was 4.0 ± 3.2 mm when an EM sensor was imbedded in the needle tip. Yet, with the EM sensor back on the needle hub, greater targeting accuracy was achieved using an US guide (3.2 ± 1.6 mm) vs. freehand (5.7 ± 3.2 mm, P = .04). For kidney, registration error was 1.8 ± 1.7 mm and respiratory motion error 4.9 ± 1.0 mm. For the deeper kidney targets, target accuracy was 4.4 ± 3.2 mm with a tip EM sensor, which was an improvement over the hub EM sensor positioning (9.3 ± 4.6 mm; P < .01). An additional source of fusion error was noted for liver. Beyond 17 ± 1 mm of respiratory motion, targets were observed to move >3 cm with US transducer/needle compression resulting in 14 ± 1.4 mm targeting accuracy.
CONCLUSIONS: A combined image-fusion/EM tracking platform can provide a high degree of needle placement accuracy (<5 mm) when targeting small lesions. Results fall within accuracy of respiratory error; with best results obtained by incorporating an EM sensor into the tip of the biopsy system.
Copyright © 2013. Published by Elsevier Inc.

Entities:  

Mesh:

Year:  2013        PMID: 23395242     DOI: 10.1016/j.acra.2012.09.020

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  11 in total

Review 1.  Navigational Tools for Interventional Radiology and Interventional Oncology Applications.

Authors:  Monzer A Chehab; Waleed Brinjikji; Alexander Copelan; Aradhana M Venkatesan
Journal:  Semin Intervent Radiol       Date:  2015-12       Impact factor: 1.513

2.  Utility of Real-time CT/MRI-US Automatic Fusion System Based on Vascular Matching in Percutaneous Radiofrequency Ablation for Hepatocellular Carcinomas: A Prospective Study.

Authors:  Seungchul Han; Jeong Min Lee; Dong Ho Lee; Jeong Hee Yoon; Won Chang
Journal:  Cardiovasc Intervent Radiol       Date:  2021-07-26       Impact factor: 2.740

3.  Clinical evaluation of a robotic system for precise CT-guided percutaneous procedures.

Authors:  Shiran Levy; S Nahum Goldberg; Ido Roth; Moran Shochat; Jacob Sosna; Isaac Leichter; Sebastian Flacke
Journal:  Abdom Radiol (NY)       Date:  2021-06-19

4.  Electromagnetic Tracking and Optical Molecular Imaging Guidance for Liver Biopsy and Point-of-Care Tissue Assessment in Phantom and Woodchuck Hepatocellular Carcinoma.

Authors:  Quirina M B de Ruiter; Sheng Xu; Ming Li; William F Pritchard; Matthew F Starost; Armando Filie; Andrew S Mikhail; Michal Mauda-Havakuk; Juan A Esparza-Trujillo; Ivane Bakhutashvili; Pedram Heidari; Umar Mahmood; John W Karanian; Bradford J Wood
Journal:  Cardiovasc Intervent Radiol       Date:  2021-05-21       Impact factor: 2.797

5.  A Novel CT to Cone-Beam CT Registration Method Enables Immediate Real-Time Intraprocedural Three-Dimensional Assessment of Ablative Treatments of Liver Malignancies.

Authors:  Marco Solbiati; Katia M Passera; S Nahum Goldberg; Alessandro Rotilio; Tiziana Ierace; Vittorio Pedicini; Dario Poretti; Luigi Solbiati
Journal:  Cardiovasc Intervent Radiol       Date:  2018-02-28       Impact factor: 2.740

6.  Multi-Operational Selective Computer-Assisted Targeting of hepatocellular carcinoma-Evaluation of a novel approach for navigated tumor ablation.

Authors:  Pascale Tinguely; Marius Schwalbe; Torsten Fuss; Dominik P Guensch; Andreas Kohler; Iris Baumgartner; Stefan Weber; Daniel Candinas
Journal:  PLoS One       Date:  2018-05-23       Impact factor: 3.240

7.  Application of Real-Time 3D Navigation System in CT-Guided Percutaneous Interventional Procedures: A Feasibility Study.

Authors:  Priya Bhattacharji; William Moore
Journal:  Radiol Res Pract       Date:  2017-10-18

8.  Electromagnetic tracking-based ultrasound/computed tomography fusion imaging in dogs: preliminary application to ocular and periocular regions.

Authors:  Jihye Choi; Sang Kwon Lee; Seungjo Park
Journal:  J Vet Sci       Date:  2019-09       Impact factor: 1.672

Review 9.  Sensor Fusion and Smart Sensor in Sports and Biomedical Applications.

Authors:  José Jair Alves Mendes; Mário Elias Marinho Vieira; Marcelo Bissi Pires; Sergio Luiz Stevan
Journal:  Sensors (Basel)       Date:  2016-09-23       Impact factor: 3.576

10.  Augmented reality for interventional oncology: proof-of-concept study of a novel high-end guidance system platform.

Authors:  Marco Solbiati; Katia M Passera; Alessandro Rotilio; Francesco Oliva; Ilaria Marre; S Nahum Goldberg; Tiziana Ierace; Luigi Solbiati
Journal:  Eur Radiol Exp       Date:  2018-07-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.