| Literature DB >> 23388431 |
Benjamin Kimble1, Kong Ming Li, Merran Govendir.
Abstract
An improved method to determine meloxicam (MEL) concentrations in koala plasma using reversed phase high performance liquid chromatography equipped with a photo diode array detector was developed and validated. A plasma sample clean-up step was carried out with hydrophilic- lipophilic copolymer solid phase extraction cartridges. MEL was separated from an endogenous interference using an isocratic mobile phase [acetonitrile and 50 mM potassium phosphate buffer (pH 2.15), 45 : 55 (v : v)] on a Nova-Pak C18 4-µm (300 × 3.9 mm) column. Retention times for MEL and piroxicam were 8.03 and 5.56 min, respectively. Peak area ratios of MEL to the internal standard (IS) were used for regression analysis of the calibration curve, which was linear from 10 to 1,000 ng/mL (r(2) > 0.9998). Average absolute recovery rates were 91% and 96% for MEL and the IS, respectively. This method had sufficient sensitivity (lower quantitation limit of 10 ng/mL), precision, accuracy, and selectivity for routine analysis of MEL in koala plasma using 250-µL sample volumes. Our technique clearly resolved the MEL peak from the complex koala plasma matrix and accurately measured MEL concentrations in small plasma volumes.Entities:
Keywords: koala; liquid chromatography; meloxicam
Mesh:
Substances:
Year: 2013 PMID: 23388431 PMCID: PMC3615235 DOI: 10.4142/jvs.2013.14.1.7
Source DB: PubMed Journal: J Vet Sci ISSN: 1229-845X Impact factor: 1.672
Fig. 1Chemical structures of meloxicam (MEL) and piroxicam (PIR).
Fig. 2(A) Chromatogram of low quality control (QC; 10 ng/mL of MEL). (B) UV spectra of MEL in the high QC sample (1,000 ng/mL of MEL). (C) Representative chromatograms of the double-blank plasma and QC samples (low concentration, 10 ng/mL; middle concentration, 200 ng/mL; high concentration, 1,000 ng/mL).
Precision and accuracy of plasma MEL QCs (triplicates per day for 5 days)
*Data are presented as the mean ± SD. CV: coefficient of variation.
Fig. 3Chromatograms of MEL in extracted from pooled koala pharmacokinetic (PK) samples [after IV injection of MEL (0.4 mg/kg)] subjected to different pretreatments for solid phase extraction (SPE). (A) Sample treated with 2% phosphoric acid prior to SPE. (B) Sample not acidified prior to SPE.
Fig. 4Semi-logarithmic graph of MEL plasma concentration vs. time in koalas (n = 3) following intravenous (IV) administration of 0.4 mg/kg MEL.