Characterization of a synthetic single crystal diamond Schottky diode for radiotherapy electron beam dosimetry.

C Di Venanzio1, Marco Marinelli, E Milani, G Prestopino, C Verona, G Verona-Rinati, M D Falco, P Bagalà, R Santoni, M Pimpinella.   

Abstract

PURPOSE: To investigate the dosimetric properties of synthetic single crystal diamond based Schottky diodes under irradiation with therapeutic electron beams from linear accelerators.
METHODS: A single crystal diamond detector was fabricated and tested under 6, 8, 10, 12, and 15 MeV electron beams. The detector performances were evaluated using three types of commercial detectors as reference dosimeters: an Advanced Markus plane parallel ionization chamber, a Semiflex cylindrical ionization chamber, and a p-type silicon detector. Preirradiation, linearity with dose, dose rate dependence, output factors, lateral field profiles, and percentage depth dose profiles were investigated and discussed.
RESULTS: During preirradiation the diamond detector signal shows a weak decrease within 0.7% with respect to the plateau value and a final signal stability of 0.1% (1σ) is observed after about 5 Gy. A good linear behavior of the detector response as a function of the delivered dose is observed with deviations below ±0.3% in the dose range from 0.02 to 10 Gy. In addition, the detector response is dose rate independent, with deviations below 0.3% in the investigated dose rate range from 0.17 to 5.45 Gy∕min. Percentage depth dose curves obtained from the diamond detector are in good agreement with the ones from the reference dosimeters. Lateral beam profile measurements show an overall good agreement among detectors, taking into account their respective geometrical features. The spatial resolution of solid state detectors is confirmed to be better than that of ionization chambers, being the one from the diamond detector comparable to that of the silicon diode. A good agreement within experimental uncertainties was also found in terms of output factor measurements between the diamond detector and reference dosimeters.
CONCLUSIONS: The observed dosimetric properties indicate that the tested diamond detector is a suitable candidate for clinical electron beam dosimetry.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23387735     DOI: 10.1118/1.4774360

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Characterization of a new commercial single crystal diamond detector for photon- and proton-beam dosimetry.

Authors:  Yuichi Akino; Archana Gautam; Len Coutinho; Jan Würfel; Indra J Das
Journal:  J Radiat Res       Date:  2015-08-12       Impact factor: 2.724

2.  Clinical radiation therapy measurements with a new commercial synthetic single crystal diamond detector.

Authors:  Wolfram U Laub; Richard Crilly
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

3.  The use of 0.5rcav as an effective point of measurement for cylindrical chambers may result in a systematic shift of electron percentage depth doses.

Authors:  Princess C Anusionwu; Jorge E Alpuche Aviles; Stephen Pistorius
Journal:  J Appl Clin Med Phys       Date:  2020-01-03       Impact factor: 2.102

4.  Design, realization, and characterization of a novel diamond detector prototype for FLASH radiotherapy dosimetry.

Authors:  Marco Marinelli; Giuseppe Felici; Federica Galante; Alessia Gasparini; Lucia Giuliano; Sophie Heinrich; Matteo Pacitti; Giuseppe Prestopino; Verdi Vanreusel; Dirk Verellen; Claudio Verona; Gianluca Verona Rinati
Journal:  Med Phys       Date:  2022-01-31       Impact factor: 4.506

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.