Validation of novel calibration scheme with traceable point-like (22)Na sources on six types of PET scanners.

Tomoyuki Hasegawa1, Keiichi Oda, Yasuhiro Wada, Toshiaki Sasaki, Yasushi Sato, Takahiro Yamada, Mikio Matsumoto, Hideo Murayama, Kei Kikuchi, Hiroki Miyatake, Yutaka Abe, Kenta Miwa, Kenta Akimoto, Kei Wagatsuma.   

Abstract

OBJECTIVE: To improve the reliability and convenience of the calibration procedure of positron emission tomography (PET) scanners, we have been developing a novel calibration path based on traceable point-like sources. When using (22)Na sources, special care should be taken to avoid the effects of 1.275-MeV γ rays accompanying β (+) decays. The purpose of this study is to validate this new calibration scheme with traceable point-like (22)Na sources on various types of PET scanners.
METHOD: Traceable point-like (22)Na sources with a spherical absorber design that assures uniform angular distribution of the emitted annihilation photons were used. The tested PET scanners included a clinical whole-body PET scanner, four types of clinical PET/CT scanners from different manufacturers, and a small-animal PET scanner. The region of interest (ROI) diameter dependence of ROI values was represented with a fitting function, which was assumed to consist of a recovery part due to spatial resolution and a quadratic background part originating from the scattered γ rays.
RESULTS: The observed ROI radius dependence was well represented with the assumed fitting function (R (2) > 0.994). The calibration factors determined using the point-like sources were consistent with those by the standard cross-calibration method within an uncertainty of ±4 %, which was reasonable considering the uncertainty in the standard cross-calibration method.
CONCLUSION: This novel calibration scheme based on the use of traceable (22)Na point-like sources was successfully validated for six types of commercial PET scanners.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23381938     DOI: 10.1007/s12149-013-0692-x

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  1 in total

1.  Evaluation of spatial dependence of point spread function-based PET reconstruction using a traceable point-like 22Na source.

Authors:  Taisuke Murata; Kenta Miwa; Noriaki Miyaji; Kei Wagatsuma; Tomoyuki Hasegawa; Keiichi Oda; Takuro Umeda; Takashi Iimori; Yoshitada Masuda; Takashi Terauchi; Mitsuru Koizumi
Journal:  EJNMMI Phys       Date:  2016-10-26
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.