Literature DB >> 23377706

A robust method to track colonoscopy videos with non-informative images.

Jianfei Liu1, Kalpathi R Subramanian, Terry S Yoo.   

Abstract

PURPOSE: Continuously, optical and virtual image alignment can significantly supplement the clinical value of colonoscopy. However, the co-alignment process is frequently interrupted by non-informative images. A video tracking framework to continuously track optical colonoscopy images was developed and tested.
METHODS: A video tracking framework with immunity to non-informative images was developed with three essential components: temporal volume flow, region flow, and incremental egomotion estimation. Temporal volume flow selects two similar images interrupted by non-informative images; region flow measures large visual motion between selected images; and incremental egomotion processing estimates significant camera motion by decomposing each large visual motion vector into a sequence of small optical flow vectors. The framework was extensively evaluated via phantom and colonoscopy image sequences. We constructed two colon-like phantoms, a straight phantom and a curved phantom, to measure actual colonoscopy motion.
RESULTS: In the straight phantom, after 48 frames were excluded, the tracking error was [Formula: see text]3 mm of 16 mm traveled. In the curved phantom, the error was [Formula: see text]4 mm of 23.88 mm traveled after 72 frames were excluded. Through evaluations with clinical sequences, the robustness of the tracking framework was demonstrated on 30 colonoscopy image sequences from 22 different patients. Four specific sequences among these were chosen to illustrate the algorithm's decreased sensitivity to (1) fluid immersion, (2) wall contact, (3) surgery-induced colon deformation, and (4) multiple non-informative image sequences.
CONCLUSION: A robust tracking framework for real-time colonoscopy was developed that facilitates continuous alignment of optical and virtual images, immune to non-informative images that enter the video stream. The system was validated in phantom testing and achieved success with clinical image sequences.

Entities:  

Mesh:

Year:  2013        PMID: 23377706     DOI: 10.1007/s11548-013-0814-x

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  18 in total

1.  Tracking of a bronchoscope using epipolar geometry analysis and intensity-based image registration of real and virtual endoscopic images.

Authors:  K Mori; D Deguchi; J Sugiyama; Y Suenaga; J Toriwaki; C R Maurer; H Takabatake; H Natori
Journal:  Med Image Anal       Date:  2002-09       Impact factor: 8.545

2.  SIFT flow: dense correspondence across scenes and its applications.

Authors:  Ce Liu; Jenny Yuen; Antonio Torralba
Journal:  IEEE Trans Pattern Anal Mach Intell       Date:  2011-05       Impact factor: 6.226

Review 3.  CT colonography reporting and data system: a consensus proposal.

Authors:  Michael E Zalis; Matthew A Barish; J Richard Choi; Abraham H Dachman; Helen M Fenlon; Joseph T Ferrucci; Seth N Glick; Andrea Laghi; Michael Macari; Elizabeth G McFarland; Martina M Morrin; Perry J Pickhardt; Jorge Soto; Judy Yee
Journal:  Radiology       Date:  2005-07       Impact factor: 11.105

4.  Nonrigid 2-D/3-D registration for patient specific bronchoscopy simulation with statistical shape modeling: phantom validation.

Authors:  Fani Deligianni; Adrian J Chung; Guang-Zhong Yang
Journal:  IEEE Trans Med Imaging       Date:  2006-11       Impact factor: 10.048

5.  Hybrid bronchoscope tracking using a magnetic tracking sensor and image registration.

Authors:  Kensaku Mori; Daisuke Deguchi; Kenta Akiyama; Takayuki Kitasaka; Calvin R Maurer; Yasuhito Suenaga; Hirotsugu Takabatake; Masaki Mori; Hiroshi Natori
Journal:  Med Image Comput Comput Assist Interv       Date:  2005

6.  Non-rigid 2D-3D registration with catheter tip EM tracking for patient specific bronchoscope simulation.

Authors:  Fani Deligianni; Adrian J Chung; Guang-Zhong Yang
Journal:  Med Image Comput Comput Assist Interv       Date:  2006

7.  Efficient and reliable schemes for nonlinear diffusion filtering.

Authors:  J Weickert; B H Romeny; M A Viergever
Journal:  IEEE Trans Image Process       Date:  1998       Impact factor: 10.856

8.  Processing differential image motion.

Authors:  J H Rieger; D T Lawton
Journal:  J Opt Soc Am A       Date:  1985-02       Impact factor: 2.129

9.  3D CT-video fusion for image-guided bronchoscopy.

Authors:  William E Higgins; James P Helferty; Kongkuo Lu; Scott A Merritt; Lav Rai; Kun-Chang Yu
Journal:  Comput Med Imaging Graph       Date:  2007-12-21       Impact factor: 4.790

10.  Computer-based System for the Virtual-Endoscopic Guidance of Bronchoscopy.

Authors:  J P Helferty; A J Sherbondy; A P Kiraly; W E Higgins
Journal:  Comput Vis Image Underst       Date:  2007-11       Impact factor: 3.876

View more
  1 in total

1.  Automated visibility map of the internal colon surface from colonoscopy video.

Authors:  Mohammad Ali Armin; Girija Chetty; Hans De Visser; Cedric Dumas; Florian Grimpen; Olivier Salvado
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-08-04       Impact factor: 2.924

  1 in total