Literature DB >> 23377578

Anterior chamber angle imaging with swept-source optical coherence tomography: detecting the scleral spur, Schwalbe's Line, and Schlemm's Canal.

Hamish McKee1, Cong Ye, Marco Yu, Shu Liu, Dennis S C Lam, Christopher K S Leung.   

Abstract

OBJECTIVE: To determine the visibility of the scleral spur (SS), Schwalbe's line (SL), and Schlemm's canal (SC) imaged by a swept-source optical coherence tomographer (OCT).
METHODS: One eye from each individual was randomly selected from 30 normal subjects and 30 patients with primary open-angle glaucoma for anterior segment imaging with a swept-source OCT. The angles were imaged with 2 protocols: (1) high-density (HD, a raster of 64 B-scans each with 512 A-scans over 8 mm) and (2) low-density scans (LD, 128 radial scans each with 512 A-scans over 16 mm). The visibility of the angle structures was determined by 2 masked observers and compared among the superior (90 degrees), nasal (0 degrees), inferior (270 degrees), and temporal (180 degrees) quadrants.
RESULTS: The interobserver agreement for assessment of visibility of the angle structures was high with agreement coefficients ranging between 0.769 and 0.987. The SS, SL, and SC were visible by both observers in 95% to 100%, 68% to 98%, and 12% to 42% of a total of 240 (4×60) quadrants in 60 HD images, respectively. The visibility was reduced in LD images (50% to 95%, 0% to 10%, and 0%, respectively). The temporal and nasal quadrants generally had superior visibility of the angle structures than the superior and inferior quadrants.
CONCLUSIONS: The SS, SL, and SC could be identified with the swept-source OCT although their visibility varied with the scan location and scan density. The SS was best visualized, followed by the SL and then the SC in the OCT images.

Entities:  

Mesh:

Year:  2013        PMID: 23377578     DOI: 10.1097/IJG.0b013e31824485fa

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  18 in total

1.  Shorter scleral spur in eyes with primary open-angle glaucoma.

Authors:  David L Swain; Joseph Ho; Julia Lai; Haiyan Gong
Journal:  Invest Ophthalmol Vis Sci       Date:  2015-02-10       Impact factor: 4.799

2.  Deep Learning Classifiers for Automated Detection of Gonioscopic Angle Closure Based on Anterior Segment OCT Images.

Authors:  Benjamin Y Xu; Michael Chiang; Shreyasi Chaudhary; Shraddha Kulkarni; Anmol A Pardeshi; Rohit Varma
Journal:  Am J Ophthalmol       Date:  2019-08-22       Impact factor: 5.258

3.  Evaluating glaucoma damage: emerging imaging technologies.

Authors:  Tigran Kostanyan; Gadi Wollstein; Joel S Schuman
Journal:  Expert Rev Ophthalmol       Date:  2015-02-09

4.  Four-dimensional microscope- integrated optical coherence tomography to enhance visualization in glaucoma surgeries.

Authors:  Neel Dave Pasricha; Paramjit Kaur Bhullar; Christine Shieh; Christian Viehland; Oscar Mijail Carrasco-Zevallos; Brenton Keller; Joseph Adam Izatt; Cynthia Ann Toth; Pratap Challa; Anthony Nanlin Kuo
Journal:  Indian J Ophthalmol       Date:  2017-01       Impact factor: 1.848

Review 5.  Intra-operative optical coherence tomography in glaucoma surgery-a systematic review.

Authors:  Bryan C H Ang; Sheng Yang Lim; Syril Dorairaj
Journal:  Eye (Lond)       Date:  2019-11-26       Impact factor: 3.775

6.  Benefit of Measuring Anterior Segment Structures Using an Increased Number of Optical Coherence Tomography Images: The Chinese American Eye Study.

Authors:  Benjamin Y Xu; Paul Israelsen; Billy X Pan; Dandan Wang; Xuejuan Jiang; Rohit Varma
Journal:  Invest Ophthalmol Vis Sci       Date:  2016-11-01       Impact factor: 4.799

7.  Age-Related Changes in Human Schlemm's Canal: An in Vivo Optical Coherence Tomography-Based Study.

Authors:  Yujin Zhao; Junyi Chen; Xiaobo Yu; Jianjiang Xu; Xinghuai Sun; Jiaxu Hong
Journal:  Front Physiol       Date:  2018-06-05       Impact factor: 4.566

8.  Reference values for anterior chamber morphometrics with swept-source optical coherence tomography in a Caucasian population.

Authors:  Hellen Cs Römkens; Henny Jm Beckers; Jan Sag Schouten; Tos Tjm Berendschot; Carroll Ab Webers
Journal:  Clin Ophthalmol       Date:  2018-02-28

9.  Repeatability and Reproducibility of Anterior Chamber Angle Measurement with Swept-Source Optical Coherence Tomography in Patients with Primary Angle Closure Suspect.

Authors:  Tong Zhao; Mengya Zhao; Qinghong Xie; Christopher W Fong; Jeremy Chen; Yingna Liu; Hui Feng; Shuning Li; Benjamin F Arnold; Tin Aung; Mingguang He; Julius T Oatts; Ying Han
Journal:  Curr Eye Res       Date:  2021-07-12       Impact factor: 2.424

Review 10.  Ageing and ocular surface immunity.

Authors:  Alireza Mashaghi; Jiaxu Hong; Sunil K Chauhan; Reza Dana
Journal:  Br J Ophthalmol       Date:  2016-07-04       Impact factor: 4.638

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.