Literature DB >> 23377070

Comparison of the accuracy of Biomet 3i Encode Robocast Technology and conventional implant impression techniques.

Kent J Howell1, Edwin A McGlumphy, Carl Drago, Gregory Knapik.   

Abstract

PURPOSE: To compare the accuracy of implant master casts fabricated using Robocast Technology (Biomet 3i) with that of master casts fabricated using traditional transfer (closed-tray) and pick-up (open-tray) techniques.
MATERIALS AND METHODS: A stereolithographic replica of a Kennedy Class I human mandible was fabricated for use as the master model. Implants were placed into both posterior quadrants (both second premolars and second molars) and set parallel (P) on one side and divergent (nonparallel, NP) on the opposite side. Impressions were made of the master model (patient replica model) with Encode Healing Abutments, open-tray, and closed-tray impression copings. Identical metallic spheres were placed onto each implant analog in the stone master casts, and the casts/spheres were scanned using a digital scanner. Measurements were made between the center points of the spheres and compared to the master model. Data were divided into P, NP, and individual sites, and the differences were analyzed statistically (α = .05).
RESULTS: Encode master casts were less accurate than the open-tray casts in NP sites. Encode master casts were less accurate than the open-tray and closed-tray casts in P sites. NP sites demonstrated less accuracy than P sites within the Encode group. Encode master casts were less accurate than the open- and closed-tray casts at the mandibular right second premolar site. The mandibular left second premolar was less accurate than the mandibular right second molar in the Encode group.
CONCLUSIONS: Within the limitations of this lab-based study and analysis, the Encode technique resulted in master casts that were less accurate than master casts made from traditional open- and closed-tray impression techniques. Further research is necessary before specific clinical judgments can be made.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23377070     DOI: 10.11607/jomi.2546

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  4 in total

Review 1.  Accuracy of Digital Dental Implants Impression Taking with Intraoral Scanners Compared with Conventional Impression Techniques: A Systematic Review of In Vitro Studies.

Authors:  María Isabel Albanchez-González; Jorge Cortés-Bretón Brinkmann; Jesús Peláez-Rico; Carlos López-Suárez; Verónica Rodríguez-Alonso; María Jesús Suárez-García
Journal:  Int J Environ Res Public Health       Date:  2022-02-11       Impact factor: 3.390

2.  Comparison of accuracy between digital and conventional implant impressions: two and three dimensional evaluations.

Authors:  Chuang Bi; Xingyu Wang; Fangfang Tian; Zhe Qu; Jiaming Zhao
Journal:  J Adv Prosthodont       Date:  2022-08-29       Impact factor: 1.989

3.  Evaluation of accuracy of various impression techniques and impression materials in recording multiple implants placed unilaterally in a partially edentulous mandible- An in vitro study.

Authors:  G Parameshwari; B Chittaranjan; N Sudhir; C-K Anulekha-Avinash; M Taruna; M Ramureddy
Journal:  J Clin Exp Dent       Date:  2018-04-01

4.  Can transfer type and implant angulation affect impression accuracy? A 3D in vitro evaluation.

Authors:  Davide Farronato; Pietro Mario Pasini; Veronica Campana; Diego Lops; Lorenzo Azzi; Mattia Manfredini
Journal:  Odontology       Date:  2021-06-01       Impact factor: 2.634

  4 in total