Effect of screw insertion torque on push-out and cantilever bending properties of five different angle-stable systems.

Alessandro Boero Baroncelli1, Ulrich Reif, Cristina Bignardi, Bruno Peirone.   

Abstract

OBJECTIVE: To compare the screw push-out strength and resistance to cantilever bending of 5 different angle-stable systems using 4 different insertion torque values to tighten locking screws. STUDY
DESIGN: In vitro mechanical testing of 5 screw-plate constructs. SAMPLE POPULATION: Screw plate constructs (n = 60) were tested; 12 of each design, 3 for each torque value.
METHODS: To compare push-out strength, screws were loaded in axial direction on the screw tip until loosening of the locking mechanism was recorded. For cantilever bending test, screws were loaded perpendicularly to their longitudinal axis at 2 mm of distance from the under surface of the plate. Load was applied in displacement control at 1 mm/min.
RESULTS: There was a significant difference between the 5 different angle-stable systems regarding both push-out and cantilever bending strength. There was an influence of insertion torque value on push-out strength for 2 systems and insertion torque value influenced cantilever bending behavior only in 1 locking system.
CONCLUSIONS: Locking mechanisms using "thread in thread" principle provided a stronger screw push-out behavior. Screws materials and core diameter of the different screws were directly related to cantilever bending strength. © Copyright 2013 by The American College of Veterinary Surgeons.

Mesh:

Year:  2013        PMID: 23373545     DOI: 10.1111/j.1532-950X.2013.01088.x

Source DB:  PubMed          Journal:  Vet Surg        ISSN: 0161-3499            Impact factor:   1.495


  3 in total

1.  Numerical Simulation of an Intramedullary Elastic Nail: Expansion Phase and Load-Bearing Behavior.

Authors:  Giulia Pascoletti; Filippo Cianetti; Giovanni Putame; Mara Terzini; Elisabetta M Zanetti
Journal:  Front Bioeng Biotechnol       Date:  2018-11-21

2.  Surgical Treatments for Canine Anterior Cruciate Ligament Rupture: Assessing Functional Recovery Through Multibody Comparative Analysis.

Authors:  Giovanni Putame; Mara Terzini; Cristina Bignardi; Brian Beale; Don Hulse; Elisabetta Zanetti; Alberto Audenino
Journal:  Front Bioeng Biotechnol       Date:  2019-08-06

3.  Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates.

Authors:  Christopher Ioannou; Matthew Knight; Luca Daniele; Lee Flueckiger; Ezekiel S L Tan
Journal:  J Orthop Surg Res       Date:  2016-10-17       Impact factor: 2.359

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.