Literature DB >> 23372057

A novel 8-mm Schlemm's canal scaffold reduces outflow resistance in a human anterior segment perfusion model.

Vikas Gulati1, Shan Fan, Cassandra L Hays, Thomas W Samuelson, Iqbal Ike K Ahmed, Carol B Toris.   

Abstract

PURPOSE: To study the effect on outflow facility and outflow resistance of a nitinol microstent implanted into Schlemm's canal.
METHODS: Using a constant pressure perfusion method, outflow facility and outflow resistance were measured in 26 pairs of dissected anterior segments from donated human eyes. Measurements were made at perfusion pressures of 10, 20, 30 and 40 mm Hg. The Hydrus Microstent was placed in Schlemm's canal of one eye and the contralateral eye underwent a sham procedure. Outflow facility and outflow resistance were measured again after the microstent implantation or sham procedure.
RESULTS: The Hydrus Microstent significantly increased outflow facility from 0.33 ± 0.17 μL/min/mm Hg to 0.52 ± 0.19 μL/min/mm Hg (P < 0.001). Outflow resistance was significantly reduced from 4.38 ± 3.03 mm Hg/μL/min at baseline to 2.34 ± 1.04 mm Hg/μL/min (P < 0.001) with the microstent. There was a linear correlation between outflow resistance at baseline and decrease in outflow resistance with the microstent (R(2) = 0.89, P < 0.0001).
CONCLUSIONS: The increase in outflow facility and decrease in resistance supports the potential use of the Hydrus Microstent as a surgical option to reduce intraocular pressure (IOP). The IOP-lowering effect may be higher in eyes with higher outflow resistance (and IOP) as compared with eyes with lower outflow resistance (and IOP).

Entities:  

Mesh:

Year:  2013        PMID: 23372057     DOI: 10.1167/iovs.12-11373

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  16 in total

1.  IOP elevation reduces Schlemm's canal cross-sectional area.

Authors:  Larry Kagemann; Bo Wang; Gadi Wollstein; Hiroshi Ishikawa; Jessica E Nevins; Zach Nadler; Ian A Sigal; Richard A Bilonick; Joel S Schuman
Journal:  Invest Ophthalmol Vis Sci       Date:  2014-03-25       Impact factor: 4.799

Review 2.  Techniques and outcomes of minimally invasive trabecular ablation and bypass surgery.

Authors:  Kevin Kaplowitz; Joel S Schuman; Nils A Loewen
Journal:  Br J Ophthalmol       Date:  2013-12-12       Impact factor: 4.638

3.  Outflow enhancement by three different ab interno trabeculectomy procedures in a porcine anterior segment model.

Authors:  Yalong Dang; Chao Wang; Priyal Shah; Susannah Waxman; Ralitsa T Loewen; Ying Hong; Hamed Esfandiari; Nils A Loewen
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2018-05-02       Impact factor: 3.117

4.  Improvement in outflow facility by two novel microinvasive glaucoma surgery implants.

Authors:  Cassandra L Hays; Vikas Gulati; Shan Fan; Thomas W Samuelson; Iqbal Ike K Ahmed; Carol B Toris
Journal:  Invest Ophthalmol Vis Sci       Date:  2014-03-28       Impact factor: 4.799

5.  One Year of Glaucoma Research in Review: 2012 to 2013.

Authors:  Charles Kim; Anna M Demetriades; Nathan M Radcliffe
Journal:  Asia Pac J Ophthalmol (Phila)       Date:  2014 Jan-Feb

Review 6.  Effectiveness and limitations of minimally invasive glaucoma surgery targeting Schlemm's canal.

Authors:  Masayuki Kasahara; Nobuyuki Shoji
Journal:  Jpn J Ophthalmol       Date:  2020-11-05       Impact factor: 2.447

Review 7.  Aqueous outflow imaging techniques and what they tell us about intraocular pressure regulation.

Authors:  Jed A Lusthaus; Tasneem Z Khatib; Paul A R Meyer; Peter McCluskey; Keith R Martin
Journal:  Eye (Lond)       Date:  2020-08-21       Impact factor: 3.775

Review 8.  Microinvasive glaucoma surgery: a review and classification of implant-dependent procedures and techniques.

Authors:  Joanna Jabłońska; Katarzyna Lewczuk; Joanna Konopińska; Zofia Mariak; Marek Rękas
Journal:  Acta Ophthalmol       Date:  2021-05-14       Impact factor: 3.988

9.  Modeling the effects of glaucoma surgery on intraocular pressure.

Authors:  Nicholas Farrar; David B Yan; Mark Johnson
Journal:  Exp Eye Res       Date:  2021-05-25       Impact factor: 3.770

Review 10.  Update on Minimally Invasive Glaucoma Surgery (MIGS) and New Implants.

Authors:  Lívia M Brandão; Matthias C Grieshaber
Journal:  J Ophthalmol       Date:  2013-11-27       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.