Literature DB >> 23369151

Study of head scatter factor in 4MV photon beam used in radiotherapy.

Murugan Appasamy1, Sidonia Valas Xavier, Thayalan Kuppusamy, Ramasubramanian Velayudham.   

Abstract

The 4 MV photon beam offers equal build-up region behavior like Co-60 beam and it plays a major role in head and neck and pediatric radiotherapy. In this study an attempt is made to study the head scatter factor (SC) for 4 MV photon beam using locally designed PMMA and Brass miniphantoms. The SC is measured in combination of PMMA miniphantom with 0.6 cc chamber and Brass miniphantom with 0.6 cc and 0.13 cc chambers. The measured SC is compared with the literature data and it agrees within ± 1.98%. The study reveals that either 0.13 cc or 0.6 cc chamber with PMMA or Brass phantom materials can be used for SC measurements in a 4 MV photon beam. The variation of SSD does not alter the head scatter factor. The collimator exchange effect is found to be within 1, and it is less than that of other linear accelerators. It is also found that the presence of internal wedge has significant contribution to head scatter factor. The Phantom scatter factor is also calculated and it agrees within ±1% with published data.

Entities:  

Mesh:

Year:  2013        PMID: 23369151     DOI: 10.7785/tcrt.2012.500316

Source DB:  PubMed          Journal:  Technol Cancer Res Treat        ISSN: 1533-0338


  1 in total

1.  Interpretation of In-air Output Ratio of Wedged Fields in Different Measurement Conditions.

Authors:  Parinaz Mehnati; Farideh Biglari; Ali Jomehzadeh
Journal:  J Med Signals Sens       Date:  2019 Apr-Jun
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.