Literature DB >> 23369050

Clinical advantages of a metal stent with an S-shaped anti-reflux valve in malignant biliary obstruction.

Kyong Joo Lee1, Moon Jae Chung, Jeong Youp Park, Don Haeng Lee, Suk Jung, Byoung Wook Bang, Seung Woo Park, Jae Bock Chung, Si Young Song, Seungmin Bang.   

Abstract

BACKGROUND AND AIM: Transpapillary stent insertion is the standard treatment for palliating malignant biliary obstruction. However, luminal occlusion often occurs due to sludge formation, tumor ingrowth, or tumor overgrowth. Currently, influx of duodenal contents by duodenobiliary reflux is considered a mechanism of stent obstruction. The aim of the present study was to evaluate the efficacy and safety of a metal stent with an anti-reflux valve in distal malignant biliary obstruction.
METHODS: Thirty-two patients diagnosed with unresectable malignant biliary obstruction were enrolled. This trial was conducted at two tertiary hospitals in Korea from June 2009 to June 2011. An uncovered metal stent (M.I.Tech) with an S-type anti-reflux valve was placed.
RESULTS: A total of 32 patients were enrolled. The technical success rate of stent insertion was 100%. The median duration of stent patency was 14.4 months. Stent occlusion occurred in 11 cases, which was caused by sludge formation in six cases, tumor ingrowth in four cases, and migration of the stent membrane in one case. During the follow-up period, there were no episodes of ascending cholangitis, acute pancreatitis, or bleeding.
CONCLUSIONS: The new metal stent with an S-shaped anti-reflux valve demonstrated a relatively long duration of stent patency. This was attributable to reductions in duodenobiliary reflux by the anti-reflux valve. Also, the stent with an S-shaped anti-reflux valve is technically feasible and very safe. However, further prospective, randomized comparison studies of stents with anti-reflux valves and conventional stents are needed to evaluate the duration of stent patency.
© 2012 The Authors. Digestive Endoscopy © 2012 Japan Gastroenterological Endoscopy Society.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23369050     DOI: 10.1111/j.1443-1661.2012.01382.x

Source DB:  PubMed          Journal:  Dig Endosc        ISSN: 0915-5635            Impact factor:   7.559


  8 in total

1.  A novel lumen-apposing metal stent with an anti-reflux valve for endoscopic ultrasound-guided drainage of pseudocysts and walled-off necrosis: A pilot study.

Authors:  In Rae Cho; Moon Jae Chung; Jung Hyun Jo; Hee Seung Lee; Jeong Youp Park; Seungmin Bang; Seung Woo Park; Si Young Song
Journal:  PLoS One       Date:  2019-09-04       Impact factor: 3.240

2.  Design and Optimization of a New Anti-reflux Biliary Stent With Retractable Bionic Valve Based on Fluid-Structure Interaction Analysis.

Authors:  Yushan Su; Zhongxia Xiang; Xiaofei Song; Shuxian Zheng; Xinsheng Xu
Journal:  Front Bioeng Biotechnol       Date:  2022-03-28

3.  Usefulness of duckbill-type anti-reflux self-expandable metal stents for distal malignant biliary obstruction with duodenal invasion: A pilot study.

Authors:  Ikuhiro Kobori; Yasumi Katayama; Fuki Hayakawa; Takeshi Fujiwara; Masaru Kuwada; Yoshinori Gyotoku; Akihiro Kitahama; Yumi Kusano; Masaya Tamano
Journal:  DEN open       Date:  2022-03-09

Review 4.  Basic Knowledge about Metal Stent Development.

Authors:  Seok Jeong
Journal:  Clin Endosc       Date:  2016-03-22

5.  Antireflux Metal Stent as a First-Line Metal Stent for Distal Malignant Biliary Obstruction: A Pilot Study.

Authors:  Tsuyoshi Hamada; Hiroyuki Isayama; Yousuke Nakai; Osamu Togawa; Naminatsu Takahara; Rie Uchino; Suguru Mizuno; Dai Mohri; Hiroshi Yagioka; Hirofumi Kogure; Saburo Matsubara; Natsuyo Yamamoto; Yukiko Ito; Minoru Tada; Kazuhiko Koike
Journal:  Gut Liver       Date:  2017-01-15       Impact factor: 4.519

6.  Antireflux Metal Stent for Initial Treatment of Malignant Distal Biliary Obstruction.

Authors:  Shinichi Morita; Yasuaki Arai; Shunsuke Sugawara; Miyuki Sone; Yasunari Sakamoto; Takuji Okusaka; Shigetaka Yoshinaga; Yutaka Saito; Shuji Terai
Journal:  Gastroenterol Res Pract       Date:  2018-01-31       Impact factor: 2.260

7.  Early malfunction of a biliary self-expandable metal stent with an antireflux valve: A case report.

Authors:  Sang Hoon Kim; Chi Hyuk Oh; Jae Min Lee; Seong Ji Choi; Hyuk Soon Choi; Eun Sun Kim; Bora Keum; Yoon Tae Jeen; Hoon Jai Chun; Hong Sik Lee; Chang Duck Kim
Journal:  Medicine (Baltimore)       Date:  2020-04       Impact factor: 1.817

8.  Evaluation of valve function in antireflux biliary metal stents.

Authors:  Chang-Il Kwon; Jong Pil Moon; Ho Yun; Seok Jeong; Dong Hee Koh; Woo Jung Lee; Kwang Hyun Ko; Dae Hwan Kang
Journal:  BMC Gastroenterol       Date:  2018-10-19       Impact factor: 3.067

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.