| Literature DB >> 23366776 |
Abstract
Micro-electrodes and micro-electrode arrays (MEAs) for stimulating neurons or recording action potentials are widely used in medical applications or biological research. For medical implants in many applications like brain implants or retinal implants there is a need for flexible MEAs with a large area and a large number of stimulation electrodes. In this work a flexible MEA with an embedded flexible silicon dummy CMOS-chip facing these challenges has been designed, manufactured and characterized. This approach offers the possibility by connecting and addressing several of these MEAs via a bus system, to increase the number and the density of electrodes significantly. This paper describes the design and fabrication process. Results on the mechanical and electrical behavior will be given and possible improvements for medical applications by this novel approach will be discussed.Entities:
Mesh:
Substances:
Year: 2012 PMID: 23366776 DOI: 10.1109/EMBC.2012.6346815
Source DB: PubMed Journal: Conf Proc IEEE Eng Med Biol Soc ISSN: 1557-170X