Literature DB >> 23358405

Quantitative evaluation on [⁹⁰Y] DOTATOC PET and SPECT imaging by phantom acquisitions and clinical applications in locoregional and systemic treatments.

C Fabbri1, V Mattone, M Casi, F De Lauro, M Agostini, N Bartolini, M D'arienzo, G Marchi, M Bartolomei, G Sarti.   

Abstract

AIM: The aim of this paper was to compare [⁹⁰Y]-PET and SPECT imaging quantification for dosimetric applications in targeted radionuclide therapy.
METHODS: Imaging studies were carried out by SPECT-CT and PET equipment performing phantom tests first. [⁹⁰Y]-SPECT and PET scans were compared in terms of sensitivity, minimum detectable activity concentration, recovery coefficients (RCs) and system spatial resolution (FWHM). Quantitative evaluations by PET and SPECT acquisitions were then assessed in patients who received therapeutic activity of [⁹⁰Y]-DOTATOC directly injected into the surgical cavity by locoregional route in glioma treatment and by systemic route in neuroendocrine tumour patients who underwent intravenous infusion. Finally 3D-dose distributions by SPECT and PET images were obtained.
RESULTS: Sensitivity was proven to be about fivefold higher for SPECT than for PET. To obtain a good-quality PET imaging, the minimum detectable activity concentration was determined to be equal to 1 MBq/mL compared with 0.05 MBq/mL that was sufficient to assess adequate SPECT imaging. RCs were 100% for volume ≥ 25.5 mL for PET and ≥ 110 mL for SPECT. FWHM was 7 mm for PET and 19 mm for SPECT scans. With regard to locoregional therapy, excellent imaging was obtained with both PET and SPECT. On the contrary, systemic administration did not permit us to obtain suitable PET imaging. PET and SPECT images were affected by considerable noise, whose influence is much more important in the quantitative evaluation of dose volume histograms rather than in the visual interpretation of images.
CONCLUSION: [⁹⁰Y]-activity quantification is feasible by SPECT and PET imaging. For clinical applications, SPECT-CT is the best technique for visualizing the radiopharmaceuticals following systemic infusion, while both SPECT and PET scans are effective in analyzing locoregional distribution. Nevertheless PET study demonstrates the best spatial definition.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23358405

Source DB:  PubMed          Journal:  Q J Nucl Med Mol Imaging        ISSN: 1824-4785            Impact factor:   2.346


  3 in total

1.  Y-90 SPECT ML image reconstruction with a new model for tissue-dependent bremsstrahlung production using CT information: a proof-of-concept study.

Authors:  Hongki Lim; Jeffrey A Fessler; Scott J Wilderman; Allen F Brooks; Yuni K Dewaraja
Journal:  Phys Med Biol       Date:  2018-05-22       Impact factor: 3.609

Review 2.  Theranostic Imaging of Yttrium-90.

Authors:  Chadwick L Wright; Jun Zhang; Michael F Tweedle; Michael V Knopp; Nathan C Hall
Journal:  Biomed Res Int       Date:  2015-05-28       Impact factor: 3.411

3.  A novel radioguided surgery technique exploiting β(-) decays.

Authors:  E Solfaroli Camillocci; G Baroni; F Bellini; V Bocci; F Collamati; M Cremonesi; E De Lucia; P Ferroli; S Fiore; C M Grana; M Marafini; I Mattei; S Morganti; G Paganelli; V Patera; L Piersanti; L Recchia; A Russomando; M Schiariti; A Sarti; A Sciubba; C Voena; R Faccini
Journal:  Sci Rep       Date:  2014-03-20       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.