PURPOSE: This study aimed to assess the feasibility and potential usefulness of a new intraoperative portable high-resolution imaging γ-camera for the localization of sentinel lymph node in patients with breast cancer.
METHODS: Sixteen T1 to T2 breast cancer female patients were evaluated (mean age, 62 years) with preoperative lymphoscintigraphy, acquired 30 minutes after intradermal injection of 60 to 70 MBq (1.6-1.9 mCi) of (99m)Tc-nanocolloids in saline volumes of 0.1 to 0.2 mL. Surgery took place the day after (18-20 hours after tracer injection). For intraoperative sentinel lymph node localization, a traditional γ-probe was used by the surgeon. Moreover, a portable imaging γ-camera was used by the nuclear physician who was present in the operating room.
RESULTS: The portable imaging γ-camera showed very high spatial resolution (2.4 mm) in the 4.4 × 4.4 cm(2) FOV, with a good sensitivity of 180 cps/MBq at a Plexiglas depth of 1 cm. In 11 patients, lymphoscintigraphy, γ-probe, and imaging γ-camera depicted the same number of radioactive lymph nodes (17 nodes; 5 of which were metastatic and detected in 6 different patients). In 6 patients, the portable imaging γ-camera detected 1 node more than lymphoscintigraphy (in total, 5 nodes more), one of which was metastatic (the only one metastatic in that patient).
CONCLUSIONS: Our preliminary results showed that a portable high-resolution hand-held imaging γ-camera is a feasible, not time-consuming, noninvasive procedure in intraoperative sentinel node localization, offering extra confidence to the surgeon. In our hands, it was a very useful auxiliary imaging tool especially in the identification of nodes located deep in the axilla, which are difficult to detect at the preoperative lymphoscintigraphy. Additional multicenter studies involving a greater number of patients are necessary to confirm these promising data.

Entities:  

Mesh:

Year:  2013        PMID: 23357823     DOI: 10.1097/RLU.0b013e31827a278d

Source DB:  PubMed          Journal:  Clin Nucl Med        ISSN: 0363-9762            Impact factor:   7.794


  3 in total

1.  First experience with a new technique: Portable gamma camera usage for sentinel lymph node identification in a patient with breast cancer.

Authors:  Zehra Pınar Koç; Pınar Pelin Özcan Kara; Mustafa Berkeşoğlu; Tamer Akça
Journal:  Turk J Surg       Date:  2018-03-01

2.  The Utility of Intraoperative Handheld Gamma Camera for Detection of Sentinel Lymph Nodes in Melanoma.

Authors:  Elgin Ozkan; Aydan Eroglu
Journal:  Nucl Med Mol Imaging       Date:  2015-05-13

Review 3.  Pelvic lymph node dissection in high-risk prostate cancer.

Authors:  Luciano Haiquel; Xavier Cathelineau; Rafael Sanchez-Salas; Petr Macek; Fernando Secin
Journal:  Int Braz J Urol       Date:  2022 Jan-Feb       Impact factor: 1.541