Impact of changes in haematocrit level and platelet count on thromboelastometry parameters.

Michael Nagler1, Silvio Kathriner, Lucas M Bachmann, Walter A Wuillemin.   

Abstract

INTRODUCTION: To what extent haematocrit levels (Hct) and platelet counts (PLT) influence the measurement of parameters of thromboelastometry when assessed with the ROTEM® device is unclear. We investigated to what extent thromboelastometry measurements depend on Hct and PLT.
MATERIALS AND METHODS: Whole blood samples were taken for in-vitro preparations of mixtures with three different levels of PLT and a varying Hct. Maximum clot firmness (MCF), clotting time (CT), clot formation time (CFT) and alpha angle (α) for INTEM, EXTEM, FIBTEM and APTEM was recorded.
RESULTS: Measurements depended substantially on Hct and PLT. MCF readings were systematically lower with increasing Hct (0.2 vs. 0.4: -7.8 (-8.3 to -7.2); p<0.001, 0.2 vs. 0.55: -14.5 (-17.3 to -14.3); p<0.001) but higher with increasing PLT (50 vs. 125×10(9)/l: 8.2 (4.2 to 12.3); p=0.005, 50 vs. 250×10(9)/l: 12.0 (7.2 to 16.8); p=0.002). CT readings were systematically higher with increasing Hct (0.2 vs. 0.4: 9.2 (6.2 to 12.1); p=0.001, 0.2 vs. 0.55: 38.2 (21.5 to 54.9); p=0.003) while increasing PLT had no influence. CFT readings were also systematically higher with increasing Hct (0.2 vs. 0.4: 83.8 (40.2 to 127.6); p=0.006, 0.2 vs. 0.55: 226.2 (110.7 to 341.7); p=0.006) but systematically lower with increasing PLT (50 vs. 125×10(9)/l: -144.0 (-272.3 to -15.6); p=0.036, 50 vs. 250×10(9)/l: -189.2 (-330.4 to -48.0); p=0.02); readings of the alpha angle showed a similar pattern.
CONCLUSIONS: Our results suggest that readings of thromboelastometry parameters need to be adjusted by Hct and PLT to avoid potential confounding and miss-interpretations in clinical practice.
Copyright © 2013 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23347682     DOI: 10.1016/j.thromres.2013.01.009

Source DB:  PubMed          Journal:  Thromb Res        ISSN: 0049-3848            Impact factor:   3.944


  7 in total

1.  Thromboelastometry and Thrombelastography Analysis under Normal Physiological Conditions - Systematic Review.

Authors:  Marcel Adler; Sandra Ivic; Nicolas S Bodmer; Hugo Ten Cate; Lucas M Bachmann; Walter A Wuillemin; Michael Nagler
Journal:  Transfus Med Hemother       Date:  2017-03-08       Impact factor: 3.747

2.  Predictive value of rotational thromboelastometry during cardiopulmonary bypass for thrombocytopenia and hypofibrinogenemia after weaning of cardiopulmonary bypass.

Authors:  Sung-Mi Ji; Sung-Hoon Kim; Jae-Sik Nam; Hye-Joo Yun; Jeong-Hyun Choi; Eun-Ho Lee; In-Cheol Choi
Journal:  Korean J Anesthesiol       Date:  2015-05-28

3.  Effects of different colloid infusions on ROTEM and Multiplate during elective brain tumour neurosurgery.

Authors:  N Li; S Statkevicius; B Asgeirsson; U Schött
Journal:  Perioper Med (Lond)       Date:  2015-09-29

4.  Thromboelastometry and Platelet Function during Acclimatization to High Altitude.

Authors:  Alistair S Rocke; Gordon G Paterson; Matthew T Barber; Alexander I R Jackson; Shona E Main; Calum Stannett; Martin F Schnopp; Martin MacInnis; J Kenneth Baillie; Elizabeth H Horn; Carl Moores; Paul Harrison; Alastair F Nimmo; A A Roger Thompson
Journal:  Thromb Haemost       Date:  2018-01-05       Impact factor: 5.249

5.  Next generation viscoelasticity assays in cardiothoracic surgery: Feasibility of the TEG6s system.

Authors:  Gabor Erdoes; Hannes Schloer; Balthasar Eberle; Michael Nagler
Journal:  PLoS One       Date:  2018-12-20       Impact factor: 3.240

6.  Expression of Tissue Factor in Epithelial Ovarian Carcinoma Is Involved in the Development of Venous Thromboembolism.

Authors:  Manabu Sakurai; Koji Matsumoto; Masahiko Gosho; Akiko Sakata; Yoshihiko Hosokawa; Yuri Tenjimbayashi; Takashi Katoh; Ayumi Shikama; Haruna Komiya; Hiroo Michikami; Nobutaka Tasaka; Azusa Akiyama-Abe; Sari Nakao; Hiroyuki Ochi; Mamiko Onuki; Takeo Minaguchi; Hiroyuki Yoshikawa; Toyomi Satoh
Journal:  Int J Gynecol Cancer       Date:  2017-01       Impact factor: 3.437

7.  Monitoring DOACs with a Novel Dielectric Microsensor: A Clinical Study.

Authors:  Debnath Maji; Aman Opneja; Michael A Suster; Kara L Bane; Brigid M Wilson; Pedram Mohseni; Evi X Stavrou
Journal:  Thromb Haemost       Date:  2020-09-02       Impact factor: 5.249

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.