Evaluation of vaporized hydrogen peroxide fumigation as a method for the bio-decontamination of the high efficiency particulate air filter unit.

Hai Quan Jia1, Yan Ju Li, Bei Sun, Si Qing Zhao, Ying Yi, Ming Zhao, Zong Xing Zhang, Xin Pan, Jian Cheng Qi.   

Abstract

OBJECTIVE: To evaluate the performance of vaporized hydrogen peroxide (VHP) for the bio-decontamination of the high efficiency particulate air (HEPA) filter unit.
METHODS: Self-made or commercially available bioindicators were placed at designated locations in the HEPA filter unit under VHP fumigation. The spores on coupons were then extracted by 0.5 h submergence in eluent followed by 200- time violent knocks.
RESULTS: Due to the presence of HEPA filter in the box, spore recovery from coupons placed at the bottom of the filter downstream was significantly higher than that from coupons placed at the other locations. The gap of decontamination efficiency between the top and the bottom of the filter downstream became narrower with the exposure time extended. The decontamination efficiency of the bottom of the filter downstream only improved gently with the injection rate of H2O2 increased and the decontamination efficiency decreased instead when the injection rate exceeded 2.5 g/min. The commercially available bioindicators were competent to indicate the disinfection efficiency of VHP for the HEPA filter unit.
CONCLUSION: This assay developed can detect all 16 β-lactams demanded by the European Union (EU). The whole procedure takes only 45 min and can detect 42 samples and the standards with duplicate analysis.
Copyright © 2013 The Editorial Board of Biomedical and Environmental Sciences. Published by China CDC. All rights reserved.

Entities:  

Keywords:  Bio-decontamination; Fumigation; High efficiency particulate air filter unit; Spore killing; Vaporized hydrogen peroxide

Mesh:

Substances:

Year:  2013        PMID: 23336134     DOI: 10.3967/0895-3988.2013.02.005

Source DB:  PubMed          Journal:  Biomed Environ Sci        ISSN: 0895-3988            Impact factor:   3.118


  4 in total

1.  Improvement of Vivarium Biodecontamination through Data-acquisition Systems and Automation.

Authors:  Shakthi Rk Devan; Suresh Vasu; Yogesha Mallikarjuna; Ramkumar Ponraj; Gireesh Kamath; Suresh Poosala
Journal:  J Am Assoc Lab Anim Sci       Date:  2018-03-01       Impact factor: 1.232

2.  Kinetics of Inactivation of Bacillus subtilis subsp. niger Spores and Staphylococcus albus on Paper by Chlorine Dioxide Gas in an Enclosed Space.

Authors:  Tao Wang; Jinhui Wu; Jiancheng Qi; Limei Hao; Ying Yi; Zongxing Zhang
Journal:  Appl Environ Microbiol       Date:  2016-05-02       Impact factor: 4.792

3.  Determining the Effectiveness of Decontamination with Ionized Hydrogen Peroxide.

Authors:  Miguel A Grimaldo; Donald H Bouyer; Claudio L Mafra de Siqueira
Journal:  Appl Biosaf       Date:  2020-09-01

4.  Plasma-Activated Aerosolized Hydrogen Peroxide (aHP) in Surface Inactivation Procedures.

Authors:  Camille Freyssenet; Stéphane Karlen
Journal:  Appl Biosaf       Date:  2019-03-01
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.