| Literature DB >> 23326653 |
Ronald J Rivera1, Dennis R Holmes, Lorraine Tafra.
Abstract
Breast conserving surgery has been accepted as the optimal local therapy for women with early breast cancer, emphasizing the necessity to balance oncologic goals with patient satisfaction and cosmetic outcomes. In the move to enhance a surgeon's ability to achieve histologically clear margins intraoperatively at the initial surgery, the MarginProbe (Dune Medical Devices, Caesarea, Israel) has emerged as an effective tool to accomplish that task. Based on previously reported success using the device, we assessed cosmesis and tissue resection volumes among participants in a randomized-controlled trial comparing the standard of care lumpectomy performed with and without the MarginProbe. The use of the MarginProbe device resulted in a 57% reduction in reexcision rates compared to the control group with a small increase in tissue volume removed at the primary lumpectomy. When total tissue volumes removed were analyzed, the device and control groups were still very similar after normalization to bra cup size. We concluded that the MarginProbe is an effective device to assist surgeons in determining margin status intraoperatively while allowing for better patient cosmetic outcomes due to the smaller volumes of tissue resected and the reduction in patient referrals for second surgeries due to positive margins.Entities:
Year: 2012 PMID: 23326653 PMCID: PMC3541558 DOI: 10.1155/2012/868623
Source DB: PubMed Journal: Int J Surg Oncol ISSN: 2090-1402
Figure 1Photograph (a) of MarginProbe device showing vacuum mechanism that ensures full contact of the 7 mm sensor with the margin surface. Schematic (b) showing 7 mm diameter sensor and radiofrequency field that penetrates margin surface to detect cancer-associated electromagnetic changes.
Figure 2Visual binary signal display showing either positive (Pos.) or negative (Neg.) margin at multiple sites measured on each specimen margin surface [deep (D), anterior (A), inferior (I), superior (S), lateral (L), and medial (M)].
Total tissue volume removed.
| Average per patient | Control | Device | Difference |
|---|---|---|---|
| Initial surgery | |||
| Main specimen | 61.3 cc | 59.7 cc | −1.6 cc |
| True positive shavings | 2.7 cc | 6.7 cc | 4.0 cc |
| False positive shavings | 7.7 cc | 21.1 cc | 13.4 cc |
|
| |||
| Total volume | 71.9 cc | 87.5 cc | 15.6 cc |
|
| |||
| Reexcision surgeries | |||
| Tissue volume | 49.5 cc | 28.4 cc | −21.1 cc |
| Normalized tissue volume | 5.6% | 4.0% | −1.6% |
|
| |||
| All surgeries | |||
| Total tissue volume | 84.8 cc | 93.3 cc | 8.5 cc |
| Normalized total tissue volume | 12.5% | 15.1% | 2.6% |