| Literature DB >> 23314352 |
Satyanarayana Achanta1, Minh Ngo, Allison Veitenheimer, Lara K Maxwell, Jarrad R Wagner.
Abstract
A highly sensitive and specific liquid chromatography-atmospheric pressure chemical ionization-mass spectrometry (LC/APCI-MS/MS) method has been developed and validated for simultaneous quantification of vinblastine and its metabolite, desacetylvinblastine, in canine plasma and urine samples. Plasma and urine samples were processed by a solid phase extraction procedure. The optimal chromatographic behavior of these analytes was achieved on pentafluorophenyl (PFP) propyl analytical column (5μm, 50×2.1mm) under isocratic elution of 0.75mL/min with a mobile phase of 5mM ammonium acetate and methanol. The samples were analyzed in positive ion, multiple reaction monitoring mode. The calibration curves were linear over 0.125-2ng/mL (lower calibration curve); 2-100ng/mL (higher calibration curve) and 0.125-5ng/mL for vinblastine and desacetylvinblastine in plasma, and over 1-2000ng/mL and 0.5-100ng/mL for vinblastine and desacetylvinblastine in urine samples, respectively. The limits of quantitation of vinblastine and desacetylvinblastine were 0.125ng/mL in both matrices. The intra and interday accuracy was above 89% and precision below 8.6% for both analytes in both matrices. The developed method was successfully applied to ongoing in vivo vinblastine pharmacokinetic studies in dogs.Entities:
Mesh:
Substances:
Year: 2012 PMID: 23314352 DOI: 10.1016/j.jchromb.2012.11.012
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205