Verification of high dose rate brachytherapy dose distributions with EBT3 Gafchromic film quality control techniques.

Antony L Palmer1, Andrew Nisbet, David Bradley.   

Abstract

It is essential that quality control (QC) techniques are developed to keep pace with modern high dose rate (HDR) brachytherapy. Current QC methods may be insufficient to fully assure the accuracy of 3D-optimized dose delivery. This work presents an evaluation of Gafchromic EBT3 film, with multi-channel analysis, in HDR dose environments for advanced QC and commissioning. 'Film-array in water' and 'three-channel Solid Water block' purpose-designed phantoms are utilized. Dose and dose-rate dependency and practical film usage has been evaluated. EBT3 measurements of dose with radial distance from a HDR source are compared to Monte Carlo data. Semi-3D dose distributions around clinical HDR applicators are compared to treatment plans. The measurement of delivery accuracy for inverse-planned pseudo-clinical test cases, with correct delivery and simulated treatment errors, has also been investigated. Local gamma criteria of 3%, 3 mm is recommended with passing rates of at least 96% typically achieved. The system is sensitive to simulated errors in HDR delivery, with significant reductions of passing rate. It has been demonstrated that EBT3 Gafchromic film, in combination with multi-channel analysis, is appropriate for applicator, treatment unit and planning system commissioning measurements as well as practical routine QC to confirm agreement of planned and delivered complex HDR brachytherapy dose distributions.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23306148     DOI: 10.1088/0031-9155/58/3/497

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  6 in total

1.  LET dependent response of GafChromic films investigated with MeV ion beams.

Authors:  V Grilj; D J Brenner
Journal:  Phys Med Biol       Date:  2018-12-18       Impact factor: 3.609

2.  Verification of high-dose-rate brachytherapy treatment planning dose distribution using liquid-filled ionization chamber array.

Authors:  A B Mohamed Yoosuf; Prakash Jeevanandam; Glenn Whitten; Geraldine Workman; Conor K McGarry
Journal:  J Contemp Brachytherapy       Date:  2018-04-30

3.  Gafchromic EBT3 film dosimetry in electron beams - energy dependence and improved film read-out.

Authors:  Petri Sipilä; Jarkko Ojala; Sampsa Kaijaluoto; Ilkka Jokelainen; Antti Kosunen
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

4.  Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning.

Authors:  Antony L Palmer; David A Bradley; Andrew Nisbet
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

5.  Real-time dosimetry with radiochromic films.

Authors:  Pierluigi Casolaro; Luigi Campajola; Giovanni Breglio; Salvatore Buontempo; Marco Consales; Andrea Cusano; Antonello Cutolo; Francesco Di Capua; Francesco Fienga; Patrizio Vaiano
Journal:  Sci Rep       Date:  2019-03-29       Impact factor: 4.379

6.  Evaluating dosimetric accuracy of the 6 MV calibration on EBT3 film in the use of Ir-192 high dose rate brachytherapy.

Authors:  Lyu Huang; Hani Gaballa; Jenghwa Chang
Journal:  J Appl Clin Med Phys       Date:  2022-02-28       Impact factor: 2.243

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.