Experimental evaluation of an electromechanical artificial urinary sphincter in an animal model.

Massimo Valerio1, Patrice Jichlinski, Roland Dahlem, Piergiorgio Tozzi, Anthony R Mundy.   

Abstract

UNLABELLED: What's known on the subject? and what does the study add?: The AMS 800 urinary control system is the gold standard for the treatment of urinary incontinence due to sphincter insufficiency. Despite excellent functional outcome and latest technological improvements, the revision rate remains significant. To overcome the shortcomings of the current device, we developed a modern electromechanical artificial urinary sphincter. The results demonstrated that this new sphincter is effective and well tolerated up to 3 months. This preliminary study represents a first step in the clinical application of novel technologies and an alternative compression mechanism to the urethra.
OBJECTIVES: To evaluate the effectiveness in continence achievement of a new electromechanical artificial urinary sphincter (emAUS) in an animal model. To assess urethral response and animal general response to short-term and mid-term activation of the emAUS.
MATERIALS AND METHODS: The principle of the emAUS is electromechanical induction of alternating compression of successive segments of the urethra by a series of cuffs activated by artificial muscles. Between February 2009 and May 2010 the emAUS was implanted in 17 sheep divided into three groups. The first phase aimed to measure bladder leak point pressure during the activation of the device. The second and third phases aimed to assess tissue response to the presence of the device after 2-9 weeks and after 3 months respectively. Histopathological and immunohistochemistry evaluation of the urethra was performed.
RESULTS: Bladder leak point pressure was measured at levels between 1091 ± 30.6 cmH2 O and 1244.1 ± 99 cmH2 O (mean ± standard deviation) depending on the number of cuffs used. At gross examination, the explanted urethra showed no sign of infection, atrophy or stricture. On microscopic examination no significant difference in structure was found between urethral structure surrounded by a cuff and control urethra. In the peripheral tissues, the implanted material elicited a chronic foreign body reaction. Apart from one case, specimens did not show significant presence of lymphocytes, polymorphonuclear leucocytes, necrosis or cell degeneration. Immunohistochemistry confirmed the absence of macrophages in the samples.
CONCLUSIONS: This animal study shows that the emAUS can provide continence. This new electronic controlled sequential alternating compression mechanism can avoid damage to urethral vascularity, at least up to 3 months after implantation. After this positive proof of concept, long-term studies are needed before clinical application could be considered.
© 2013 BJU International.

Entities:  

Keywords:  artificial urinary sphincter; intrinsic sphincter deficiency; stress urinary incontinence

Mesh:

Year:  2013        PMID: 23305222     DOI: 10.1111/j.1464-410X.2012.11728.x

Source DB:  PubMed          Journal:  BJU Int        ISSN: 1464-4096            Impact factor:   5.588


  3 in total

Review 1.  Review of surgical implant procedures for male incontinence after radical prostatectomy according to IDEAL framework.

Authors:  Dimitri Barski; Holger Gerullis; Thomas Otto
Journal:  Updates Surg       Date:  2017-05-06

2.  Montreal electronic artificial urinary sphincters: Our futuristic alternatives to the AMS800™.

Authors:  Xavier Biardeau; Sami Hached; Oleg Loutochin; Lysanne Campeau; Mohamad Sawan; Jacques Corcos
Journal:  Can Urol Assoc J       Date:  2017-10       Impact factor: 1.862

3.  A novel urethra compression technique using Turkish continence device for male urinary incontinence

Authors:  Öner Odabaş; Irem Gül Sancak; Yusuf Kasap; Zerrin Mahmut; Erkan Ölçücüoğlu; Sedat Taştemur; Neslihan Zengin
Journal:  Turk J Med Sci       Date:  2020-08-26       Impact factor: 0.973

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.